emission condenser Peltronic Manual

Table Of Contents
page 51 of 120
Catchpot at outlet:
Remove joint clip. Remove catchpot and drain conden-
sate.
Catchpot at inlet:
Admit air or inert gas (via the pump inlet) to restore at-
mospheric pressure in the catchpot before attempting
removal. Remove joint clip. Remove catchpot and drain
condensate.
Reattach drained catchpots.
+ Important: Comply with regulations when disposing
of solvents/condensates. Recycle if possible; purify if
contaminated.
Shutdown & storage
The pump can be switched off under vacuum.
Short-term:
Has the pump been exposed to condensate?
- Allow the pump to continue to run at atmospheric pres-
sure for a few minutes.
Has the pump been exposed to media which may damage
the pump materials or form deposits?
- Check and clean pump heads if necessary.
Has the pressure transducer been exposed to media which
may form deposits?
- Clean pressure transducer at the CVC 3000 controller
if necessary.
Long-term:
- Take measures as described above regarding short-
term shutdown.
- Separate the pump from the application.
- Close inlet and outlet ports (e.g., with transport caps).
- Close the gas ballast valve.
- Drain catchpots.
- Store the pump under dry conditions.
NOTICE
NOTICE