emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 62 of 120
A preset ”Duration” (process time) has no effect if the process is
stopped due to a preset ”Maximum” before ”Duration” is reached.
The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be
set to ”Off”. ”Off” means that no endpoint of the process is dened.
The screen-shots show the factory-set values.
When selecting ”Graphic” the display shows a pressure vs. time curve.
The timeline in the diagram adapts automatically to the process time.
+ Press the selection knob twice to return to the standard display.
Temporarily switching from ”Vac control” to ”Auto mode” while pro-
cess control is running:
+ Press ”MODE” key. The controller switches to ”Auto mode” and adapts
the boiling pressure starting with the actual set value. The preset func-
tion of the controller does not change due to this switching. When
pressing the ”STOP” key, the controller is in ”Vac control” function
again.
Adjustment of the set vacuum during vacuum control:
Dynamic, interactive adaptation:
+ Press the selection knob and keep pressed.
+ Turning the knob for a 1/4 turn to the left causes pump down.
+ Turning the knob for a 1/4 turn to the right causes venting.
+ When the knob is released, the current pressure value is used as new
set value.
Vac control
Vac control
Set vacuum 75 Torr
Speed HI
Maximum Off
Delay Off
Duration Off
- - - - - - Graphic - - - - - -
- - - - - - - Back - - - - - - -
760 Torr