emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 66 of 120
Program function
Function denition:
➨ Permits ten programs to be dened and stored, each with up to
ten program steps with preset values for vacuum and time.
+ Edit:
Use to dene the preset values for the process run:
Time: Denes either the process runtime for each program step to
reach a preset vacuum level or, if programming a ”Step”, the runtime
after having achieved the vacuum level. The total process runtime is
shown in the base line. Attention: A preset runtime of 99:59:59 hours
in the nal program step will cause the process to run endlessly. Ter-
minate the process by pressing the ”STOP” key .
Vacuum: Vacuum level to be attained.
Venting valve: Operating a venting valve to reach a preset vacuum
level.
”Step”: ”Step” causes pump down as fast as possible to the preset
vacuum level. As soon as the vacuum level is reached the time meter
starts running.
”Auto”: Auto =
➜
: indicates that the system will search for a boiling
point automatically, and will automatically adapt to changes in the boil-
ing point until a preselected time interval expires.
Auto =
: indicates that the system will automatically adapt to chang-
es in the boiling point. The program step is nished when the selected
time or pressure value has been reached.
Selecting ”Auto” in combination with ”Step” in one program step is not
possible.
+ Open: Choose program (Programs 0 - 9).
+ Store: This command stores an edited program or the program of the
last process to one of the storage spaces 0 - 9.
+ Delay: ”Delay” species the time the coolant valve remains open after
process stop.
The ”Delay” is adjustable in a range of 1-300 minutes or can be set to
”Off” (”Off” means that the coolant valve closes immediately when the
process stops.).