emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 69 of 120
VACUULAN function
Function denition:
➨ Optimizes vacuum control for vacuum networks
(e.g., VACUUBRAND VACUU•LAN)
➨ Permits on-demand operation of a speed-controlled pump
(VACUU•BUS-compatible VARIO pump)
Preselections
+ Use the selection knob to select the parameters.
+ Set vacuum (the lower switch-off value): If the pressure drops below
the ”Set vacuum”, a time-meter starts to run and the motor speed is re-
duced. When the pressure exceeds the ”Set vacuum” pressure again,
the time meter is reset. If the ”Set vacuum” is not reached within 100
hours, the controller signals an error.
The ”Set vacuum” is adjustable in the range of 1-795 Torr (1-1060
mbar).
+ Switch on (the higher switching value): If the pressure exceeds this
pressure, pumping down starts again.
The ”Switch on” pressure is adjustable in the range of 2-795 Torr (2-
1060 mbar).
In the event of a sudden high-pressure spike, pumping starts again
even if the ”Switch on” pressure has not been reached (pressure in-
crease control).
+ Delay: If the vacuum is below ”Set vacuum” for longer than the ”Delay”
time, the pump is stopped. Pumping starts again in the event of a rapid
pressure increase or if the ”Switch on” level is exceeded.
The ”Delay” is adjustable in a range of 1-300 minutes or can be set to
”Off” (”Off” means that the pump stops immediately when the pressure
drops below ”Set vacuum”.).