emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 71 of 120
Assembly of a vacuum system
+ Assemble the vacuum connection line between the vacuum pump and
the vacuum application.
+ Assemble the vacuum connection line between the controller and the
vacuum application, if the controller is not part of a pumping unit.
+ Assemble electrical connections.
+ Connect coolant if necessary.
Vacuum for ltration and suction
+ Select ”Pump down” function.
+ If necessary, set value for ”Speed” (high or low pumping speed). Re-
duce the speed in case of fast sedimentation (formation of a lter cake).
+ Set ”Minimum” to a value which provides adequate suction but does
not lead to evaporation of the solvent.
+ Start process by pressing ”START/STOP” key.
alternatively:
+ Select ”Vac control” function.
+ Set ”Set vacuum” to a value which does not lead to evaporation of the
solvent.
+ If necessary, set value for ”Speed” (high or low pumping speed). Re-
duce the speed in case of fast sedimentation (formation of a lter cake).
+ Set ”Maximum” pressure setting so that pump will switch off at the end
of the ltration process, or in the event that a lter cake cracks, leading
to a sudden pressure increase.
+ Start process by pressing ”START/STOP” key.
+ Tip for ltration: Adjust preset pressure to a value well above the
boiling pressure of the solvent (e.g., for water >>15 Torr (20 mbar)).
Set the maximum pressure to e.g., 375 Torr (500 mbar). Once the l-
tration has nished, the pressure increases and the pump is switched
off automatically.
Application examples