emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 72 of 120
+ Select function ”Pump down” function. ”Speed” ”HI” is recommended.
For gel dryers set a lower speed if necessary (if the gels tend to crack).
+ Set ”Minimum” to prevent volatile components from evaporating. The
process is stopped automatically as soon as ”Minimum” is reached.
+ Set a process time (”Duration”) if necessary.
+ Start process by pressing ”START/STOP” key.
alternatively:
+ Select ”Vac control” function to dry at a predetermined vacuum level.
+ Set ”Set vacuum” to the preferred evaporation vacuum of the solvent.
+ Set a process time (”Duration”) if necessary.
+ Start process by pressing ”START/STOP” key.
Vacuum for gel dryer, drying chambers and
vacuum concentrators
Fully automatic determination and adaptation of the boiling point
with a VACUU•BUS-compatible VARIO pump (recommended)
+ Select ”Auto mode”.
+ Start process by pressing ”START/STOP” key.
+ The ”Auto mode” allows a fully automatic distillation even if the boiling
vacuum is unknown. With solvent mixtures, the vacuum is reduced
until all solvents or volatile components are evaporated (with respect
to the ultimate vacuum of the pump and the bath temperature).
+ If ”Minimum” is set, the controller switches off the pump when the pre-
set vacuum value is reached. The ”Minimum” should be well below
the boiling vacuum of the solvent to be evaporated and above the ul-
timate vacuum attainable in the apparatus. If the ”Minimum” is set to
”Auto” , the controller automatically switches off the pump at the end
of the evaporation (recommended for large solvent quantities only).
The setting of ”Duration” (process runtime) has no effect if the pump is
switched off due to ”Minimum” before ”Duration” is over.
Vacuum for distillation and evaporation
(e.g., rotary evaporator)