emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 73 of 120
+ When setting a value for ”Duration” the controller switches off the
pump when ”Duration” has passed even if a preset ”Minimum” is still
not reached.
+ If neither ”Minimum” nor ”Duration” is preset, pumping down has to be
nished by pressing the ”START/STOP” key.
alternatively:
Semi-automatic distillation and evaporation
+ Select function ”Pump down”.
+ Start process by pressing ”START/STOP” key.
+ Observe process. As soon as evaporation starts, press ”MODE” key
(switching to ”Vac control”). The vacuum level is kept constant (at the
boiling pressure). Fine tuning of the vacuum value is possible by turn-
ing the selection knob.
Pressing ”MODE” key a second time switches to ”Auto mode” causing
the vacuum level to be automatically adapted to changes of the boiling
pressure.
alternatively:
+ Select ”Vac control” function.
+ Set a maximum ”Speed”, to select a faster or slower process.
+ To set ”Maximum” is usually not necessary, unlike ltrations, because
the pressure does not increase at the end of the evaporation.
+ Set a value for ”Duration” if the process should be terminated auto-
matically after a denite time.
+ Set ”Set vacuum” depending on the solvent and the temperature.
+ Start process by pressing ”START” key.
+ To temporarily switch to ”Auto mode”: Press ”MODE” key. The
controller switches to ”Auto mode”. The setting of the controller does
not change due to this switching. After pressing ”START/STOP” key
the controller is in ”Vac control” again.