emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 75 of 120
Conguration
In the ”Conguration” menu the device parameters are preselected.
After 20 seconds without action the function ”Conguration” and its sub-
menus (except submenu ”Sensors”) are quit automatically without stor-
ing any possibly changed parameter.
+ Adjustment: Adjustment of the pressure transducer under vacuum
and/or at atmospheric pressure, see also section „Readjustment of
CVC 3000“, pg. 80.
Adjustment to atmospheric pressure is carried out at an absolute pres-
sure value between 795 - 525 Torr (1060 - 700 mbar). This is espe-
cially helpful in high elevation laboratories. Adjustment under vacuum
may be done at an absolute vacuum value between 0 - 15 Torr (0 - 20
mbar). In the range between 15 to 525 Torr (20 to 700 mbar) no adjust-
ment is possible; ---- Torr is displayed.
+ RS-232: Conguration of the interface, setting of parameters and com-
mands, see section ”Interface”.
Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”,
”7-O-1” or ”7-E-1”, Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” and
remote on ”On” or ”Off”.
+ Sensors: Selection of the pressure transducer to be controlled (maxi-
mum 4 external sensors).
+ Display: Selection of the device parameters ”Brightness” between 0
- 100%, ”Contrast” between 0 - 100%, ”Sound” ”On” or ”Off”, ”Units”
”mbar”, ”hPa” or ”Torr”, ”Language” ”German”, ”English”, ”French”,
”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”,
”Russian”, ”Polish”, ”Dutch”, ”Japanese”, ”Finnish”.
+ Autostart: If ”Autostart” is set to ”On” the controller restarts a running
process automatically after a power failure. If this is unwanted, set
”Autostart” to ”Off”.
Preselections
+ Use the selection knob to select the parameters.