emission condenser Peltronic Manual

Table Of Contents
page 75 of 120
Conguration
In the ”Conguration” menu the device parameters are preselected.
After 20 seconds without action the function ”Conguration” and its sub-
menus (except submenu ”Sensors”) are quit automatically without stor-
ing any possibly changed parameter.
+ Adjustment: Adjustment of the pressure transducer under vacuum
and/or at atmospheric pressure, see also section „Readjustment of
CVC 3000“, pg. 80.
Adjustment to atmospheric pressure is carried out at an absolute pres-
sure value between 795 - 525 Torr (1060 - 700 mbar). This is espe-
cially helpful in high elevation laboratories. Adjustment under vacuum
may be done at an absolute vacuum value between 0 - 15 Torr (0 - 20
mbar). In the range between 15 to 525 Torr (20 to 700 mbar) no adjust-
ment is possible; ---- Torr is displayed.
+ RS-232: Conguration of the interface, setting of parameters and com-
mands, see section ”Interface”.
Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”,
”7-O-1” or ”7-E-1”, Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” and
remote on ”On” or ”Off”.
+ Sensors: Selection of the pressure transducer to be controlled (maxi-
mum 4 external sensors).
+ Display: Selection of the device parameters ”Brightness” between 0
- 100%, ”Contrast” between 0 - 100%, ”Sound” ”On” or ”Off”, ”Units
”mbar”, ”hPa” or ”Torr”, ”Language” ”German”, ”English”, ”French”,
”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”,
”Russian”, ”Polish”, ”Dutch”, ”Japanese”, ”Finnish”.
+ Autostart: If ”Autostart” is set to ”On” the controller restarts a running
process automatically after a power failure. If this is unwanted, set
Autostart” to ”Off”.
Preselections
+ Use the selection knob to select the parameters.