emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 87 of 120
Write commands ”CVC 2000”
Command Operation Parameter Description
OUT_MODE function
1
2
3
30
31
32
continuous pumping
vacuum control without automatic
vacuum control with automatic
optional: sensitivity: low
optional: sensitivity: normal
optional: sensitivity: high
OUT_SP_1 set vacuum XXXX
unit (mbar/Torr/hPa) according to pre-
selection; see respective function for
parameter range
OUT_SP_V
set vacuum
with venting*
XXXX
unit (mbar/Torr/hPa) according to pre-
selection; see respective function for
parameter range
OUT_SP_2 set frequency XX.X
motor speed in Hz
(99.9 for “HI”; 99.8 for 3000 rpm)
OUT_SP_3
vacuum for switch
on (VACUU•LAN)
XXXX
unit (mbar/Torr/hPa) according to pre-
selection; see respective function for
parameter range
OUT_SP_4 delay XX:XX hh:mm (hours:minutes)
OUT_SP_5
vacuum for auto-
matic switching
off
XXXX
unit (mbar/Torr/hPa) according to pre-
selection; see respective function for
parameter range
OUT_SP_6
time for automatic
switching off
(VACUU•LAN)
XX:XX hh:mm (hours:minutes)
START
starting process
control
STOP
stopping process
control
1
2
termination of process control
termination of process control and stor-
age of the current pressure as new set
point
REMOTE
remote
operation**
0
1
set controller to local operation
set controller to remote operation
OUT_VENT
driving venting
valve
0
1
close venting valve (not automatically)
open venting valve (process control
stopped)
* Pressure setting with venting is only possible in ”Vac control” function
if a venting valve is connected and congured, and vacuum control is
started. The venting valve opens automatically if the actual pressure
is at least 7.5 Torr (10 mbar) below the preset pressure. Automatic
venting becomes inactive if vacuum control is stopped (by pressing
”START/STOP” or ”VENT”), a pressure value is set using the com-
mand OUT_SP_1, or if the function is changed. Activate the command
OUT_SP_V again if necessary.