emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 91 of 120
Write commands ”CVC 3000”
Command Operation Parameter Description
OUT_MODE function
0
1
2
3
30
31
32
4
VACUU•LAN
Pump down
Vac control
Auto mode
optional: sensitivity: low
optional: sensitivity: normal
optional: sensitivity: high
Program
OUT_CFG
conguration
(bus monitoring)
yXXX
X0XX
X1XX
X2XX
XX0X
XX1X
XXX0
XXX1
y: 0......D: language
+
(hexadecimal), see
”Read commands CVC 3000”
pressure unit mbar
pressure unit Torr
pressure unit hPa
Autostart off
Autostart on
acoustic signal off
acoustic signal on
Attention (OUT_MODE): If control is running, it is only possible to switch either from 1 to 2, or
from 2 to 3, or from 3 to 2. The set vacuum is adopted in each case.
Command Operation Response Description
IN_SP_4 delay XX:XX h:m hours:minutes (00:00 = Off)
IN_SP_5
switch off
pressure
XXXX mbar/Torr/hPa
”Maximum” for ”Vac control”,
”Minimum” for ”Pump down”)
unit according to preselections
IN_SP_6 runtime XX:XX h:m process runtime (hours:minutes)
IN_SP_P1y time XX:XX:XX h:m:s
time in program step y (0......9)
(hours:minutes:seconds)
IN_SP_P2y pressure XXXX mbar/Torr/hPa
pressure in program step y (0......9)
unit according to preselections
IN_SP_P3y venting valve
0
1
no venting valve in program step y
(0......9)
venting valve in program step y
(0.....9)
IN_SP_P4y Step
0
1
no ”Step” in program step y (0.....9)
”Step” in program step y (0......9)
IN_SP_P5y Auto
0
1
no ”Auto” in program step y (0.....9)
”Auto” in program step y (0......9)
IN_PV_Sx
current pressure
of pressure trans-
ducer x
XXXX mbar/hPa/Torr
pressure of pressure transducer x
(order of numbering according to
display in ”Sensors” menu)
IN_VER version CVC 3000 VX.XX software version