emission condenser Peltronic Manual
Table Of Contents
- Part I
- Reset / Language selection
- Safety information!
- Technical data
- Use and operation
- Installing a pump in a vacuum system
- Notes regarding the (dry) ice condenser
- Notes concerning the operation of the inlet condenser IK
- Notes concerning the operation of the Peltronic emission condenser
- Notes concerning the operation of pumping units with silencer
- During operation
- Important notes regarding the use of gas ballast
- Important notes concerning the operation of the exhaust waste vapor condenser
- Shutdown & storage
- CVC 3000 Vacuum controller
- Menu guide
- Pump down function
- Vac Control function
- Auto mode
- Program function
- VACUULAN function
- Application examples
- Configuration
- Part II
- Readjustment of CVC 3000
- Calibration in the factory
- Cleaning the pressure transducer
- Interface parameters
- Accessories
- Troubleshooting
- Replacing diaphragms and valves
- Cleaning and inspecting the pump heads
- Disassembling the housing cover at the side of the emission condenser
- Replacing the diaphragm
- Assembling the housing cover at the side of the emission condenser
- Disassembling the housing cover at the side of the ON/OFF switch
- Assembling the housing cover at the side of the ON/OFF switch
- Assembling the fittings
- Cleaning and replacing components
- Notes on return to the factory
- Warranty
- Health and safety clearance form
- EC Declaration of Conformity of the Machinery
page 92 of 120
Command Operation Parameter Description
OUT-SP_1 set vacuum XXXX
unit according to preselection; see re-
spective function for parameter range
OUT_SP_V
set vacuum with
venting
XXXX
unit according to preselection; see re-
spective function for parameter range
OUT_SP_2 speed XXX speed in % or ”HI”
OUT_SP_3 start-up pressure XXXX
unit according to preselection; see re-
spective function for parameter range
OUT_SP_4 delay XX:XX hh:mm (hours:minutes)
OUT_SP_5
switch-off
pressure
XXXX
unit according to preselection; see re-
spective function for parameter range
OUT_SP_6 switch-off time XX:XX hh:mm (hours:minutes)
OUT_SP_PL open program X program 0......9
OUT_SP_PS store program X program 0......9
OUT_SP_P1y time
XX:XX:XX
+XX:XX:XX
total runtime until program step y
(0.....9) or time for program step y
(0.....9) (additive)
OUT_SP_P2y pressure XXXX
pressure at program step y (0......9)
unit according to preselection
OUT_SP_P3y venting valve
0
1
no venting valve in program step y (0...9)
venting valve in program step y (0......9)
OUT_SP_P4y Step
0
1
no ”Step” in program step y (0.....9)
”Step” in program step y (0......9)
OUT_SP_P5y Auto
0
1
2
no ”Auto” in program step y (0.....9)
”Auto ” in program step y (0......9)
”Auto ” in program step y (0......9)
START started
STOP
0
1
2
Stop and delete fault
Stop
Stop and adopt the set vacuum
REMOTE*
0
1
Remote off
Remote on
ECHO**
0
1
Echo off
Echo on,
write command with return value
CVC
2
3
CVC 2000 commands
CVC 3000 commands***
OUT_VENT
0
1
2
venting valve closed
venting valve open
venting until atmospheric pressure
(788 Torr (1050 mbar) at maximum)
STORE
store settings permanently,
if ”ECHO = 1” after realization
➜