User guide
page 16 of 102
• Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, installation in higher altitudes, conductive
pollution or external condensation on the pump.
• Do not operate this product near ames.
To the best of our knowledge the equipment is in com-
pliance with the requirements of the applicable EC-direc-
tives and harmonized standards (see ”Declaration of Con-
formity”) with regard to design, type and model. Directive
EN 61010-1 gives in detail the conditions under which the
equipment can be operated safely (see also IP degree of
protection, “Technical data”, pg. 25).
Operating conditions
➨ These pumps are not approved for operation in po-
tentially explosive atmospheres. Do not operate the
pumps in potentially explosive atmospheres.
➨ Pumps without the ”
`
” mark on the rating plate are
not approved for the pumping of potentially explo-
sive atmospheres. Do not pump potentially explo-
sive atmospheres with those pumps.
➨ Pumps bearing the ”
`
” mark on their rating plates
are approved for the pumping of potentially explo-
sive atmospheres according to their classication
II 3G IIC T3 X according to ATEX, but they are not
approved for operation in potentially explosive at-
mospheres (see section “
`
Important information:
Equipment marking (ATEX)”, pg. 23).
➨ The pumps are not suitable to pump any of the sub-
stances listed below.
Do not pump:
- unstable substances
- substances which react explosively under impact
(mechanical stress) without air
NOTICE