Technical data
Type PC 500 LAN NT PC 600 LAN NT
Maximum pumping speed*
50/60 Hz (ISO 21360)
cfm
(m
3
/h)
1.2 / 1.4
(2.0 / 2.3)
2.0 / 2.2
(3.4 / 3.8)
Ultimate vacuum (absolute)
without gas ballast
Torr
(mbar)
5.3
(7)
1.1
(1.5)
Ultimate vacuum (absolute) with
gas ballast
Torr
(mbar)
9.0
(12)
2.3
(3)
Maximum permissible inlet
pressure (absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet
pressure (absolute)
psi
(bar)
16
(1.1)
Maximum pressure
difference between inlet and
outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure
(absolute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient
temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric
moisture during operation (no
condensation)
% 30 to 85
Maximum permissible installation
altitude above mean sea level
ft
(m)
6500
(2000)
Rated motor power
hp
(kW)
0.24
(0.18)
0.34
(0.25)
No-load speed 50/60 Hz rpm 1500 / 1800
Maximum permissible range
of supply voltage ( ±10% )
Attention: Observe specications
of rating plate!
100-115 V~ 50/60 Hz, 120 V~ 60 Hz
230 V~ 50/60 Hz
Maximum rated current at:
100-115 V~ 50/60 Hz,
120 V~ 60 Hz
230 V~ 50/60 Hz
A
A
3.4
1.8
5.7
3.0
Device fuse 2 slow blow fuses 6.3A
Motor protection thermal cutout, manual reset
* Pumping speed of diaphragm pump