Installing a pump in a vacuum system
If dangerous or polluting uids could be released at
the outlet, install an appropriate system to catch and
dispose of those uids.
+ Connect a gas-tight exhaust line at the pump outlet if
necessary. Always vent exhaust gases appropriately
(e.g., into a fume hood).
+ Never block the gas outlet. The exhaust line must al-
ways be free of obstructions (no back pressure) to en-
sure an unimpeded discharge of gas. The cross-sec-
tion of the outlet tubing must be at least the size of the
pump’s exhaust connection.
+ Maximum permissible pressure: 21.8 psi (1.5 bar) ab-
solute.
+ Attention: At pressures above approximately 795 Torr
(1060 mbar) the pressure reading becomes incorrect
due to saturation of the pressure transducer. The dis-
play will ash. Release pressure immediately! Risk of
bursting!
+ Particles and dust must not be aspirated. If necessary,
you must install appropriate lters. You must ensure
their suitability concerning gas ow, chemical resis-
tance and resistance to clogging prior to use.
Use and operation
When switching on the controller VNC 2 for the very rst time, a menu
to select the language of the controller menu is displayed. Select the de-
sired language (e.g., ”English”) by pressing the t or s keys and press
”Enter” to conrm. Then select the pressure unit (”mbar”, ”Torr” or ”hPa”)
in the same way.
It is possible to access the language selection menu at any time via
the Device menu.