Fault Possible cause Remedy
No display.
Electrical power cord
not plugged in, electri-
cal supply failure?
Plug in power cord.
Check fuse.
Device fuse blown?
Identify cause of failure.
Replace device fuse at
switch output of VNC 2 or
at diaphragm pump.
Device switched off?
Switch on controller and/
or pump.
Other than above men-
tioned causes?
Contact local distributor.
Display disappears,
internal overload
protection activated.
Too much load (e.g.,
valves) connected?
Check current draw of
the connected devices
(pump, valves).
Thermal overload at
controller, ambient tem-
perature too high?
Ensure sufcient ventila-
tion.
Short circuit at connect-
ed valves?
Replace valves.
Short circuit at the
RS 232 plug?
Check plug and cable.
Other than above men-
tioned causes?
Contact local distributor.
Pressure reading
incorrect.
Pressure transducer
decalibrated?
Readjust VNC 2.
Humidity in the mea-
surement chamber?
Let the pressure trans-
ducer dry, e.g., by pump-
ing. Readjust if neces-
sary. Determine and
eliminate the cause for
humidity.
Pressure transducer
contaminated?
See ”Cleaning the pres-
sure transducer”.
Other than above men-
tioned causes?
Contact local distributor.
Error message
ErrValve”, two
blips*.
External venting valve
removed or defective?
Connect or check valve or
replace with a new one or
recongure without valve.
Switch VNC 2 off/on.
Troubleshooting