User guide

page 79 of 102
Fault Possible cause Remedy
Controller does
not respond when
pressing keys (ex-
cept ON/OFF). ”P”
is displayed.
Controller in remote
mode?
Control VNC 2 controller
via interface or switch off
remote mode.
Controller does
not respond when
operating any keys.
No change after
switching off/on.
Contact local distributor.
Pump does not
work.
Pump too hot?
Allow pump to cool down,
identify and eliminate
cause of failure. Manual
reset is necessary. Switch
off pump or unplug.
Electrical power cord
not plugged in, electri-
cal supply failure?
Plug in power cord.
Check fuse.
Pump does not
start or stops im-
mediately.
Motor overloaded?
Allow motor to cool down,
identify and eliminate
cause of failure. Manual
reset is necessary. Switch
off pump or unplug.
Overpressure in outlet
line?
Remove blockage in line,
open valve.
Pump has been ex-
posed to condensate?
Perform maintenance.
Pump does not
achieve its ultimate
vacuum or usual
pumping speed.
Outgassing substances
or vapor generated in
the process?
Check process parame-
ters.
Leak in the tubing or in
the vacuum system?
Check connections, pipe-
line and vacuum system.
Wrong setting at con-
troller?
Select mode ”continu-
ous pumping” and check
again.
Long, narrow vacuum
line?
Use lines with larger di-
ameter, length as short as
possible.
Pump has been ex-
posed to condensate?
Allow pump to run for
some minutes with atmo-
spheric pressure at the
inlet to purge.