If the pump does not achieve the ultimate vacuum:
- Whenever the diaphragms and valves have been replaced, a break-in
period of several hours is required before the pump achieves its ulti-
mate vacuum.
- In case of an unusual noise, switch off pump immediately and check
clamping disc positions.
If the specied ultimate vacuum is not achieved, and if this does not
change after the break-in period:
Check hose connectors at pump heads for leaks. If necessary recheck
valve seats and diaphragms.
Slip the PTFE tube onto the hose connec-
tion of the vacuum connection at the con-
troller. Loosen the union nut at the vacuum
connection of the controller.
Secure with union nut. Tighten the union nut
rmly nger-tight.