Round bottom flask 500 ml, coated .......................... 638497
O-ring 28 x 2,5 ......................................................... 3120553
(FPM) at the spherical ground joint of the round bottom flask at
the inlet
alternative:
O-ring 28 x 2,5 ........................................................... 638304
(FPM covered with FEP)
Remove union nut at the condenser and remove hose from
the inlet of the condenser.
Pull off old overpressure safety relief device and install new
one.
Connect hose to the inlet of the exhaust waste vapour con-
denser and tighten union nut.
Overpressure safety relief device ............................. 638821
(at the exhaust waste vapour condenser)
Cleaning and assembling components
Cleaning the pressure transducer
The controller itself is maintenance-free.
If the vacuum system is contaminated (oil, particles, etc.), contamination of the pressure transducer will
influence the accuracy of measurement.
Attention: Never use hard objects to clean the pressure transducer!
Fill the chamber of the pressure transducer with a solvent (e. g. benzene) and allow sufficient cleaning
time. Observe all regulations concerning usage and disposal of solvents!
Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.
Rinse the chamber of the pressure transducer several times with alcohol in order to remove all solvent
residues.
Never use a spiky or sharp-edged tool to clean the pressure transducer.
Allow the pressure transducer to dry.
Readjust the pressure transducer if necessary.
Tighten union nut first by hand and then tighten by one full
turn using open-ended wrench.