Operation
51BA-N°: 999279/17.10.2014
Menu
control parameter
Program
Meaning
Menu
control parameter
VACUU•LAN
Meaning
Parameter – Program
Parameter Meaning
Current program Edit program, up to 10 programs are
storable for CVC 3000.
... load from prog. Loadtheselectedprogram.
... save as prog. Save the program under the selected number.
Parameter – VACUU•LAN
Parameter Meaning
Set vacuum
(mbar)
Adjustment of the vacuum set point.
Adjustmentrange:1‑1060
Switch on
(mbar)
Maximum limit for vacuum level. If pressure
exceeds this limit, the pump begins to pump down.
Adjustmentrange:0‑1060
Delay(min)
Period for which the pump will attempt to
maintain“Setvacuum”level.Ifnopressure
increases are detected before the time period
“Delay”expires,vacuumlevelwillbeallowed
toriseto“Switchon”level.
Adjustment range: 0 - 300 minutes