User`s guide

USER'S GUIDE____________________________________________________________________
104 _________________________________________________________________ M210855EN-D
Analog Output Quantities
AMODE/ASEL
Use the serial line to select and scale the analog output quantities:
1. Connect the barometer to the PC.
2. Open the terminal connection between your PC and the
barometer.
3. Check the analog output modes with the AMODE command.
Example:
>amode
Ch1 output : 0...1 V
Ch1 slot : 4
Ch1 type : T compensated
>
4. Select and scale the quantities for the analog outputs with the
command
ASEL. Note that the optional quantities can be
selected only if they have been selected when ordering the
device.
ASEL [xxx]
where:
xxx = Quantity of channel 1
Enter always all the quantities for all the outputs. For quantities and
their abbreviations see Table 2 on page 15.
Exam
ple:
>asel P
Ch1 P low : 900.00 hPa
Ch1 P high : 1100.00 hPa
>