Table 19 Adjustment and Calibration Commands for
Barometer Module P1
Command Function
LCP1 ? Displays current linear adjustment
status
LCP1 ON/OFF Enables/Disables the linear
adjustments
LCP1 Enter linear adjustments
MPCP1 ON/OFF Enables/Disables the multipoint
adjustments
MPCP1 Enters multipoint adjustments
CDATE Shows or sets the calibration date
CTEXT Shows or sets the calibration info
text.
Opening and Closing the Adjustment Mode
Before you can proceed with the adjustment and calibration, you must
open the adjustment mode by following the instructions below. The
adjustment button with indicator LED is located in the upper corner
on the left-hand side of the motherboard. See the position of the
adjustment button in Figure 2 on page 16. To open and close the
adjustm
ent mode:
1. Open the barom
eter cover.
2. Press the
ADJ button. The adjustment mode is now available.
The red indicator LED is on.
3. Press the
ADJ button again to lock the adjustment mode. The
red indicator LED is off.
The adjustment menu on the display/keypad is visible only when the
ADJ button (on the motherboard inside the transmitter) is pressed.
Figure 44 Adjustment Menu