User`s guide

Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 87
System Information Commands
?
Use the serial line command ? to check the current barometer
configuration. Command
?? is similar but can also be used if the
barometer is in
POLL mode.
Example:
>?
PTB330 / 1.01
Serial number : A1234567
Batch number : B8901234
Output format : P #T P1 #T P2 #T DP12 #T QFE #RN
Adjust. date : 2007-05-03
Adjust. info : VAISALA
Date : 2007-05-08
Time : 13:42:01
Start mode : STOP
Baud P D S : 4800 E 7 1
Output interval: 1 s
Address : 0
Echo : ON
Module 1 : BARO-1
Module 2 : BARO-1
Module 3 : BARO-1
Module 4 : EMPTY
>
ERRS
Use the command ERRS to display all current device errors,
unacknowledged errors, and error log contents.
An error is active if the cause for the error is still present when you
execute the ERRS command. If the cause disappears, the error
message changes to the unacknowledged state. An unacknowledged
error is an error message that has not yet been displayed to the user.
Unacknowledged errors are automatically cleared when they are
displayed with
ERRS command or with the optional display and
keypad.