Using Serial Line
NOTE
You can use the serial
commands to change/view the user port settings
even if you are currently connected to the service port.
SERI
Use the SERI command to set the communication settings for the user
port. Changed settings will be activated at next reset or power up.
SERI [b p d s]<cr>
where
b
=
Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)
p
=
Parity (n = none, e = even, o = odd)
d
=
Data bits (7 or 8)
s
=
Stop bits (1 or 2)
The settings can be changed one parameter at a time or all parameters at
once.
Example (changing all parameters):
>SERI 600 N 8 1
600 N 8 1
>
Example (changing parity only):
>SERI O
4800 O 7 1