Owner's manual

Table Of Contents
User's Guide _______________________________________________________________________
112 __________________________________________________________________ M210762EN-G
SMODE
Use the SMODE command to set the user port start-up operating mode.
SMODE [xxxx]<cr>
where
xxx
=
STOP, RUN, POLL, or MODBUS.
Table 30 Selection of Output Modes
Mode
Measurement Output
Available Commands
STOP
Only with the SEND command.
All (default mode).
RUN
Automatic output.
Only command S.
POLL
Only with the SEND [addr]
command.
SEND [addr] and OPEN [addr].
Other commands available after
opening a line to the transmitter
using the OPEN command.
Use with RS-485 buses where
multiple transmitters can share
the same line.
MODBUS
Must be read from the
transmitter using MODBUS
protocol.
Modbus protocol only; see
Chapter 5, Modbus, on page 137.
Selected output mode will be activated at next reset or power up.
ADDR
Use the ADDR command to set the device address of the transmitter.
Addresses are required for POLL mode and MODBUS mode (serial
Modbus).
ADDR [aa]<cr>
where
aa
=
Device address of the transmitter, range 0 ... 255 (default = 0)
Example (changing the transmitter address from 0 to 52):
>addr
Address : 0 ? 52
>