Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 119
Example:
0 ... 5 V voltage output selected for channel 1 and 4 ... 20 mA selected
for channel 2.
OFF
ON
Selection
1
█
2
█
Voltage output selected
3
█
4
█
5
█
6
█
0 ... 5 V selected
7
█
8
█
1
█
Current output selected
2
█
3
█
4
█
4 ... 20 mA selected
5
█
6
█
7
█
8
█
NOTE
If you have customized the error output setting (
AERR), check that the
set error values are still valid after changing the output
mode/range. See
section
Analog Output Fault Indication Setting on page 122.
Analog Output Quantities
NOTE
The relative humidity and temperature readings are not those of the
process itself but the ones measured after the cooling process.
Therefore, they should not be selected as output quantities. They are
meant to be used in calibra
tion or during installation.
Use the display/keypad to change and scale the analog output quantities.
1. Open the Main Menu by pressing any of the ▲▼◄► arrow
buttons.
2. Select Interfaces by pressing the ► arrow button.
3. Select Analog outputs by pressing the ► arrow button.
4. Select Output 1/2/3 by pressing the ► arrow button.