Owner's manual

Table Of Contents
Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 135
Example (no data logger module installed):
>pur
Interval Purge : ON ?
Interval : 1440 min ?
Power-up Purge : ON ?
Duration : 60 s ?
Settling : 240 s ?
Temperature : 180 'C ?
Temp. diff. : 0.5 'C ?
Trigger Purge : OFF ?
RH trigger : 20 %RH ?
>
Example (with data logger module installed):
>pur
Interval Purge : ON ?
Interval : 1440 min ?
Next Purge date: 2011-03-31 ?
Next Purge time: 12:00:00 ?
Power-up Purge : ON ?
Duration : 60 s ?
Settling : 240 s ?
Temperature : 180 'C ?
Temp. diff. : 0.5 'C ?
Trigger Purge : OFF ?
RH trigger : 20 %RH ?
>
NOTE
To activate
the new interval settings immediately, reset the transmitter.
NOTE
When sensor purge in power
-up is enabled, wait about 6 min after power
up before taking measurements. The output channels are locked for the
first operation minutes to the initial
measured values
Sensor Warming
DMT345 and DMT346 contain a sensor warming function, that is, when
humidity reaches a set limit (default: 80%RH), sensor warming operation
is activated automatically. Sensor warming continues as long as the
humidity is above that limit.
When the warming starts, the sensor operation indicator is displayed on
the optional display and relays configured as "online status" (if any) will
be released. Warming does not affect T
d/f
,T
d/f atm,
T
d
,T
d atm,
x, H
2
O and P
w
quantities, all others will be frozen.