Appendix A ______________________________________________ Example Installation of DMT346
VAISALA _______________________________________________________________________ 171
APPENDIX A
EXAMPLE INSTALLATION OF DMT346
This Appendix illustrates an example installation of DMT346 transmitter
into a process.
0605-045
Figure 82 Cooling Set Installation
1. Make a round 89.5 + 0.5 mm hole on the process wall.
2. Weld the tube of the mounting flange tightly on the inner metal
plate of the process wall. If the process wall is more than 125 mm
thick, a lengthening piece (max. 50 mm) can be welded on the
mounting tube (for walls thicker than 175 mm, see Figure 23 on
page 38).
3. The cooling set is mounted in a vertical position. Tighten the
screws properly in order to ensure a thermal contact.