use the floating point registers to write values outside this 16-bit signed
integer range. Maximum writable value is 32767 also for configuration
registers that do not accept negative values.
Measurement Data (Read-Only)
Table 48 Measurement Data Registers
Name
Float
Integer
Unit
RH
0001…0002
0257 (×0.01)
%
T
0003…0004
0258 (×0.01)
°C
T
d
0007…0008
0260 (×0.01)
°C
T
d/f
0009…0010
0261 (×0.01)
°C
T
d/f
(atm)
0011…0012
0262 (×0.01)
°C
T
d
(atm)
0013…0014
0263 (×0.01)
°C
a
0015…0016
0264 (×0.01)
g/m³
x
0017…0018
0265 (×0.01)
g/kg
T
w
0019…0020
0266 (×0.01)
°C
H
2
O
0021…0022
0267 (×1)
ppm
V
p
w
0023…0024
0268 (×0.1)
hPa
p
ws
0025…0026
0269 (×0.1)
hPa
h
0027…0028
0270 (×0.01)
kJ/kg
ΔT
0031…0032
0272 (×0.01)
°C
a
NTP
0033…0034
0273 (×0.01)
g/m³
SSR
0037…0038
0275 (×0.01)
%
T
s
0039…0040
0276 (×0.01)
°C
H
2
O
0065…0066
0289 (×1)
ppm
W
Available measurements depend on the instrument configuration. Values
may be unavailable also in case of device failure. Read status registers or
exception status outputs to check for failures.
NOTE
Only real-time data can be read from the instrument. Data that is
recorded in the transmitter’s memory (internal memory or data logger
module) cannot be read through Modbus.