Getting Started
Within a few seconds after power-up the LED on the cover of the
transmitter is lit continuously indicating normal operation. When the
transmitter is turned on the first time, the language selection window
opens: Select the language with ▲▼ arrow buttons and press the
SELECT button (the left-hand button).
Pressure has an effect on humidity calculations and accuracy. Therefore,
accurate calculations can be achieved only when the ambient pressure is
taken into consideration. See section Pressure Compensation Setting on
page 100 for instructions on how to set the pressure.
Start-up time for DMT345/346 transmitter is in total about 6 minutes.
The outputs (serial and analog) are activated 3 seconds after powering up
the transmitter. If the data logger module is installed, the it may take up
to 18 seconds for the outputs to be activated.
In addition, 10 seconds after the measurement the outputs will freeze for
about 6 minutes due to the sensor self diagnostics procedure (Sensor
Purge and AutoCal). The frozen output value will be the value the
transmitter reached during the 10 seconds of measurement. After the self
diagnostics procedure the outputs are operational again.