Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 93
Getting Measurement Message from
Serial Line
Starting Continuous Outputting
R
Enter the R command to start the continuous output of measurements.
Example:
>r
Tdf=-20.6 'C H2O= 958 ppmV x= 0.6 g/kg
>r
Tdf=-20.7 'C H2O= 958 ppmV T= 23.8 'C RH= 3.3 %RH
>
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars '*'.
You can change the format of the output with the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with the FORM command.
Stopping Continuous Outputting
S
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
See command SMODE to change the default (power-up) operation
mode.