Owner's manual
Table Of Contents
- DMT345/346 USER'S GUIDE
- TABLE OF CONTENTS
- GENERAL INFORMATION
- PRODUCT OVERVIEW
- INSTALLATION
- OPERATION
- Getting Started
- Display/Keypad
- MI70 Link Program for Data Handling
- Serial Line Communication
- LAN Communication
- Communicating with a Transmitter in POLL Mode
- General Settings
- Serial Output Settings
- Data Recording
- Analog Output Settings
- Operation of Relays
- Sensor Functions
- MODBUS
- MAINTENANCE
- CALIBRATION AND ADJUSTMENT
- TECHNICAL DATA
- EXAMPLE INSTALLATION OF DMT346
- CALCULATION FORMULAS
- MODBUS REFERENCE

User's Guide _______________________________________________________________________
96 ___________________________________________________________________ M210762EN-G
Formatting Serial Line Message
NOTE
Instead of using the FTIME, FDATE and FST commands described in
this section, you can use the FORM command with modifiers TIME,
DATE, and STAT. See section FORM on page 98.
FTIME and FDATE
FTIME and FDATE commands will enable/disable output of time and
date to the serial line. To add time to R and SEND outputs enter:
FTIME [x]<cr>
To add date to R and SEND outputs enter:
FDATE [x]<cr>
where
x
=
ON or OFF
Example:
>send
RH= 16.2 %RH T= 22.0 'C
>ftime on
Form. time : ON
>send
00:03:56 RH= 16.2 %RH T= 22.0 'C
>fdate on
Form. date : ON
>send
2000-01-01 00:04:08 RH= 16.2 %RH T= 22.0 'C
>
FST
Use the FST command to include the status of purge, sensor warming
and AutoCal in output from the SEND and R commands.
FST [x]<cr>
where
x
=
ON or OFF (default)