Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 97
Example:
>fst on
Form. status : ON
>send
w 0 Tdf= 0.6 'C x= 4.0 g/kg SSR= 22.5 % Ts= 23.2
'C
>
where the state of the probe is indicated by the following letters
and
values:
w...xxx
=
Sensor warming
where xxx
=
Sensor heat power
(0 when not warming)
H...xxx
=
Purge
where xxx
=
Sensor temperature (ºC)
S...xxx
=
Sensor cooling
after purge
where xxx
=
Sensor temperature (ºC)
A...xxx
=
AutoCal
where xxx
=
Sensor temperature (ºC)
General Settings
Changing Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 5 on page 20 for available quantities.
NOTE
Only the
quantities selected when ordering the device can be selected as
an output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Open the Main Menu by pressing any of the ▲▼◄► arrow
buttons.
2. Press the ► arrow button to select Display.
3. Press ► arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 4 display
quantities at a time.
5. Press EXIT to return to the basic display.
To select display units:
1. Open the Main Menu by pressing any of the ▲▼◄► arrow
buttons.