Owner's manual

Table Of Contents
User's Guide _______________________________________________________________________
98 ___________________________________________________________________ M210762EN-G
2. Press the arrow button to select Display.
3. Use the ▲▼arrow buttons to select Units. Confirm the selection
by pressing the right-hand arrow button.
4. Use the ▲▼ arrow buttons to select display units. Confirm the
selection by pressing CHANGE. The unit changes from metric to
non-metric or the other way round.
5. Press EXIT to return to the basic display.
NOTE
Changing the units by using the
display/keypad has no effect on the
serial output units.
Using Serial Line
Use the serial line command FORM to change the format, and the
command UNIT to select metric or non-metric output units.
FORM
Use the serial line command FORM to change the format or select a
certain quantities for the output commands SEND and R.
FORM [x]<cr>
where
x
=
Formatter string
Formatter string consists of quantities and modifiers. If no formatter
string is entered, the command will display the currently active formatter
string. Note that the hash symbol "#" is shown as a backslash "\" when
viewing the current formatter string.
When selecting the quantity, use the abbreviations of the quantities
presented in Table 5 on page 20 when issuing the command. The
modifiers are presented in Table 27 on page 99.