User`s guide

USER'S GUIDE____________________________________________________________________
132 _________________________________________________________________ M210762EN-D
Opening and Closing Adjustment Mode
Open the transmitter cover. The buttons needed in the adjustment are on
the left-hand side of the motherboard, see Figure 2 on page 23.
Press the ADJ button to enable the adjustment mode. The indicator led
indicates the adjustment availability.
Table 31 Indicator Led Functions
Indicator Led Function Description
LED off adjustment locked
LED on adjustment available
LED blinking evenly measurement not stabilized
LED blinking with short pulses performing Purge/AutoCal
Press the ADJ button again to disable the adjustment mode.
NOTE
Fixed pressure compensation of 1013.25 hPa is used when in adjustment
mode. Adjustments should be performed at ambient pressures.
Adjustment Information
These data are shown on the device information. (see section Device
Information on page 98. Use the display/keypad to feed the adjustment
information.
1. If you are not in the adjustment menu, press the ADJ button on the
motherboard to open the ADJUSTMENT MENU.
2. Press the arrow button to select Adjustment info.
3. Select Date, press SET. Enter the date by using the arrow buttons.
Press OK.
4. Select i, press SET. Enter information text including 17 characters
at maximum by using the arrow buttons. Press OK.
5. Press EXIT to return to the basic display.
CTEXT and CDATE
Use the serial line command CTEXT to enter text in to the adjustment
information field. First press the adjustment button on the motherboard
inside the transmitter.