Data Filtering
The averaging data filter calculates a average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Table 29 Filtering Levels
Setting Filtering level
OFF No filtering
ON (default) Standard = short filtering (about 15 s moving average)
EXTENDED Extended filtering (default: 1 min average)
Use display/keypad to set the filtering level.
1. Open the MAIN MENU by pressing any of the ▲▼◄► arrow
buttons.
2. Select Measuring by pressing the arrow button.
3. Select Filtering and press CHANGE to confirm your selection.
4. Select Off/Standard/Extended and press SELECT to confirm
your selection.
5. Press EXIT to return to the basic display.
FILT
Set the filtering level.
Syntax: FILT [xxx]
where
xxx = OFF, ON or EXT (default = ON)
Device Information
Use the display/keypad or the serial line to display the device
information.
Press the INFO button in the basic display to see the following
information:
- current sensor operation (for example, AutoCal or Purge) in progress
- present or past unacknowledged errors
- device information
- adjustment information fed by the user
- measuring settings