Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 33
In STOP mode: Measurements output only by command, all commands
can be used (default mode).
In RUN mode: Outputting automatically, only command S can be used.
In POLL mode: Measurements output only with command SEND. For
more details on POLL mode, see sectionNetworking Operation on page 37.
The settings are not valid until the device is reset.
Setting Communication Hardware
Communication hardware can be chosen between RS-232 or RS-485.
Command parameters are correspondingly 232 and 485.
RSMODE <cr>
SAVE <cr>
The setting is not valid until the device is reset.
>rs mode
RSMODE : 232? 485
>save
EEPROM saved successfully.
>