Manual

Chapter 5 _______________________________________________________________ Calibration
Calibration Table
Table 5 Greenspan's Calibration Table with Output Values
According to the Chosen Scale
Temperature
LiCl
°C
°F
%RH
15
59
*)
20
68
11.3
25
77
11.3
30
86
11.3
35
95
11.3
4 ... 20 mA mA 5.81 5.81 5.81 5.81
0 ... 20 mA mA 2.26 2.26 2.26 2.26
0 ... 1 V V 0.113 0.113 0.113 0.113
0 ... 5 V V 0.565 0.565 0.565 0.565
0 ... 10 V V 1.13 1.13 1.13 1.13
Temperature
NaCl
°C
°F
%RH
15
59
75.6
20
68
75.5
25
77
75.3
30
86
75.1
35
95
74.9
4 ... 20 mA mA 16.10 16.08 16.05 16.02 15.98
0 ... 20 mA mA 15.12 15.10 15.06 15.02 14.98
0 ... 1 V V 0.756 0.755 0.753 0.751 0.749
0 ... 5 V V 3.780 3.775 3.765 3.755 3.745
0 ... 10 V V 7.56 7.55 7.53 7.51 7.49
*) If the LiCl solution is used or stored in temperatures below +18 °C (+64 °F), the equilibrium
humidity of the salt solution may change permanently.
VAISALA _______________________________________________________________________ 25