Owner's manual

HMP228
M210282en-A Operating Manual
25
4.4.1 Calibration procedure after sensor change
Humidity calibration should be performed in stable conditions using saturated
salt solutions as a reference.
Disconnect the security lock jumper!
4.4.1.1 Using serial commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.
Insert the sensor head into the calibration hole of the LiCl chamber of
the humidity calibrator.
Wait for 30 minutes.
Give command FCRH <cr>, enter the first point value and press <cr>:
>FCRH <cr>
RH: 11.9 Ref1 ? yy.y <cr>
Press any key when ready...
The stabilization of the sensor to the reference humidity can be
monitored by giving c <cr>:
RH: 11.9 Ref1 ? c <cr>
RH: 11.5 Ref1 ? c <cr>
RH: 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
Insert the sensor head into the calibration hole of the NaCl chamber of
the humidity calibrator.
Wait for 30 minutes.
Press any key, enter the second point value and press <cr>:
RH: 75.5 Ref2 ? yy.y <cr>
The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.
4.4.1.2 Using display/keypad commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.
Insert the sensor head into the calibration hole of the LiCl chamber of
the humidity calibrator.
Wait for 30 minutes.