Owner's manual
HMP228 Transmitters
Appendix 1: Serial commands M210282en-A
48
The output scales can also be given directly with the ASCL command.
ASCL aaa.a bbb.b ccc.c ddd.d <cr>
aaa.a = lower limit of channel 1
bbb.b = upper limit of channel 1
ccc.c = lower limit of channel 2
ddd.d = upper limit of channel 2
For example, water activity is scaled to 0...1.00 on channel 1 and temperature
to -40...+100 °C on channel 2:
>ASCL 0 1 -40 100 <cr>
Ch1 (aw) lo 0.000
Ch1 (aw) hi 1.000
Ch2 (T ) lo -40.000 'C
Ch2 (T ) hi 100.000 'C