Owner's manual

HMP228
Appendix 2: Display commands M210282en-A
70
1.1.2.1 Setting the serial interface operation mode
Select Mode in the main menu; the following is displayed:
Select Serial output:
The currently valid setting flashes. Select the desired mode with the
arrow keys and press ENT. After this the programme returns to the
Mode Menu.
When Run mode is selected, the currently valid output interval is dis-
played:
The output interval setting can be changed as follows:
press CL
the number starts flashing
if the interval needs to be changed, press CL again and enter the new
interval; otherwise press ENT
the unit (s, min, h) starts flashing
the unit can be changed with the arrow keys and acknowledged with
ENT
after this the programme returns to Mode menu