Instruction Manual

HMP230 SERIES
M210225en-B Appendix 1: Serial commands
101
TIME Setting the time
TIME <cr>
For example, to enter a new time:
>TIME <cr>
Current time is 01:35:54
Enter new time (hh:mm:ss) : 13:25:56 <cr>
When the current time is asked, the new time is passed with <cr>.
VERS Name and version of the programme
VERS <cr>
For example:
>VERS <cr>
HMP 230 / x.yy
where x.yy is the programme version.
? Outputting the transmitter settings
? <cr>
For example:
>? <cr>
HMP 230 / 3.01
CPU serial nr : S4330008
Keyboard type : 1
Address : 0
Output units : metric
BaudPDS : 4800E71FDX
Serial mode : STOP
Output intrv. : 0 s
Mtim : 32
Pressure : 1013.25
Analog outputs
Ch1 0.00 ... 20.00 mA
Ch2 0.00 ... 20.00 mA
Ch1 ( RH ) lo 0.000 %RH
Ch1 ( RH ) hi 100.000 %RH
Ch2 ( T ) lo -40.000 'C
Ch2 ( T ) hi 180.000 'C
Transducer :
PRB serial nr : 9702
PRB cal. date : 970117