Instruction Manual

HMP230 SERIES
Appendix 5: Error messages M210225en-B
130
1.2 Errors during operation
Two types of errors are possible during operation of the transmitters. The first
type indicates that no frequency comes from the converter.
Display: Serial bus:
E40 f (all) out of range
- the measurement hybrid is probably damaged. Send the transmitter to
Vaisala for repair.
E41 f (T) out of range
- the temperature measurement cannot be completed. Check the temperature
sensor and its solderings. Check the temperature measurement wires and their
solderings on the component board (TP5,TP6,TP7 and TP8). If you cannot
find any faults, send the transmitter to Vaisala for repair.
E42 f (T2) out of range
- the temperature measurement cannot be completed. Check the temperature
sensor and its solderings. Check the temperature measurement wires and their
solderings on the component board (TP5,TP6,TP7 and TP8). If you cannot
find any faults, send the transmitter to Vaisala for repair.
E43 f (Rk1 ) out of range
- the temperature measurement cannot be completed. Check the temperature
sensor and its solderings. Check the temperature measurement wires and their
solderings on the component board (TP5,TP6,TP7 and TP8). If you cannot
find any faults, send the transmitter to Vaisala for repair.