• If you want to see how the sensor stabilizes to the reference
temperature, enter c <cr> instead of the first reference:
T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...
• After giving the correct temperature value (Ref1) and pressing <cr>,
press any key and press <cr>.
7.2.1.2. With display/keypad commands
• Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.
Disconnect the security lock jumper!
• Check the transmitter against the reference.
• Select Cali in the main menu and then T cal; select one-point
calibration T 1 point cal.
• Change the reading with the arrow keys to correspond to the reference
and press ENT.
7.2.1.3. With LED commands
• Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.
Disconnect the security lock jumper!
• Check the transmitter against the reference.
• Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command ¡¡l¡. At the first calibration point the LED
on the left flashes; adjust the first point (offset) with the arrow
switches to the same reading with the reference and press ENT
switch.
• After adjusting the offset point and pressing ENT the second LED
from left flashes. Press ENT without changing the output value.