If you want to see how the sensor stabilizes to the reference
temperature, enter c <cr> instead of the first reference:
T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...
After giving the correct temperature value (Ref1) and pressing <cr>,
press any key and press <cr>.
7.2.1.2. With display/keypad commands
Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.
Disconnect the security lock jumper!
Check the transmitter against the reference.
Select Cali in the main menu and then T cal; select one-point
calibration T 1 point cal.
Change the reading with the arrow keys to correspond to the reference
and press ENT.
7.2.1.3. With LED commands
Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.
Disconnect the security lock jumper!
Check the transmitter against the reference.
Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command ¡¡l¡. At the first calibration point the LED
on the left flashes; adjust the first point (offset) with the arrow
switches to the same reading with the reference and press ENT
switch.
After adjusting the offset point and pressing ENT the second LED
from left flashes. Press ENT without changing the output value.