Owner's manual

User's Guide ______________________________________________________________________
46 __________________________________________________________________ M210701EN-D
0505-247
Figure 18 Accuracy of Temperature Measurement
Table 7 Dewpoint Temperature (Calculated) Measurement
Specifications
Property Description / Value
Measuring range -20 ... +80 °C (+4 ... 176 °F)
Accuracy see Table 8 below
Table 8 Accuracy of Dewpoint Temperature Measurement
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-20 2.6 1.7 1.4 1.2 1.1 1.1 1.0 1.0
-10 2.7 1.7 1.3 1.2 1.1 1.0 1.0 1.0 0.9
0 2.3 1.4 1.1 1.0 0.9 0.8 0.8 0.8 0.7 0.7
10 2.3 1.4 1.1 0.9 0.8 0.9 0.8 0.8 0.8 0.8
20 2.2 1.2 1.0 0.8 0.7 0.6 0.6 0.5 0.5 0.6
40 3.0 1.9 1.5 1.3 1.2 1.1 1.0 1.0 1.0 1.0
60 4.3 2.7 2.2 1.9 1.8 1.7 1.6 1.5 1.5 1.5
80 4.9 3.1 2.5 2.2 2.1 2.0 1.9 1.8 1.8 1.8
Table 9 Operating Environment Specifications
Property Description / Value
Operating temperature range
Transmitter body, no display
Transmitter body, with display
Sensor head
-40 ... +60 °C (-40 ... +140 °F)
-30 ... +60 °C (-22 ... +140 °F)
-40 ... +80 °C (-40 ... +176 °F)
Storage temperature range -40 ... +60 °C (-40 ... +140 °F)
Electromagnetic compatibility EN61326-1:1197+Am1:1998; +
Am2:2001; Generic Environment