User guide

Appendix A _________________________________ Probe Installation Kits and Installation Examples
VAISALA _______________________________________________________________________ 191
Example of Installation Through Roof
Outdoor
environment
Process:
For example
+85 ÂșC, 85 %RH
0507-015
Figure 101 Example of Installation through Roof
The following numbers refer to Figure 101:
1
=
Insulated probe cables
2
=
Sealings
3
=
Roof
4
=
Cable gland for temperature probe (for example: AGRO 1100.12.91.065)
5
=
Temperature probe
6
=
Relative humidity probe
7
=
Cable gland for relative humidity probe (for example: AGRO 1160.20.145)
8
=
Plastic adapter to protect probes from condensation water coming from the
pipe. Diameter slightly smaller than tube diameter.
9
=
Plastic tube for probe (2 pcs)
10
=
Stainless steel tube coming through the roof
11
=
Two thread bars holding the plastic adapter
12
=
Insulated pipe ending
1
2
3
4
5
6
7
8
9
10
11
12