INTV
Use the INTV command to set the RUN mode output interval.
INTV [n xxx]<cr>
where
n = 1 - 255
xxx = S, MIN or H
This command sets the RUN mode output interval. The time interval is
used only when the RUN mode is active. For example, the output interval
is set to 10 minutes.
>INTV 10 min
Output intrv. : 10 min
>
Setting RUN output interval to zero enables the fastest possible output
rate.
ADDR
Addresses are required only for POLL mode (see serial line command
SMODE on page 51). Use the ADDR command to input the RS-485
transmitter address.
OPEN [aa]<cr>
where
aa
= address (0 ... 99) (default = 0)
Example: the transmitter is configured to address 99.
>ADDR
Address : 2 ? 99
>
SEND
Use the SEND command to output the reading once in POLL mode:
SEND [aa]<cr>
where
aa = address of the transmitter