Configuration Guide to MAWS TECHNICAL REFERENCE M210511EN-A November 2005
PUBLISHED BY Vaisala Oyj P.O. Box 26 FIN-00421 Helsinki Finland Phone (int.): +358 9 8949 1 Fax: +358 9 8949 2227 Visit our Internet pages at http://www.vaisala.com/ © Vaisala 2005 No part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder. The contents are subject to change without prior notice.
_________________________________________________________________________________ Table of Contents CHAPTER 1 GENERAL INFORMATION............................................................................ 5 About This Manual ................................................................... 5 Contents of This Manual ....................................................... 5 Version Information ............................................................... 6 Related Manuals ..............................
Technical Reference_________________________________________________________________ Configuring Accumulator .....................................................38 Resetting Accumulator Manually.........................................40 Using the Accumulator Variable in Reports.........................41 Power Control Option ............................................................41 Configuring Power Control Option.......................................42 Timer Option ................................
_________________________________________________________________________________ CHAPTER 7 CONFIGURING SERIAL SENSOR INTERFACES.................................... 103 Physical Interface................................................................. 104 Data Acquisition Methods ................................................... 105 Configuring Serial Line Parameters ................................... 106 Configuring Sensor Interface..............................................
Technical Reference_________________________________________________________________ This page intentionally left blank.
Chapter 1 _________________________________________________________ General Information CHAPTER 1 GENERAL INFORMATION About This Manual This manual is intended for the users of MAWS Automatic Weather Station. It provides information for creating MAWS setups with more advanced features of MAWS Lizard and is thus supplementary to MAWS Lizard Setup Software User's Guide and other MAWS user's guides.
Technical Reference_________________________________________________________________ - Chapter 5, Configuring Calculations and Reports, provides additional information on configuring different MAWS Lizard calculations and reports. - Chapter 6, Configuring Sensors, provides advanced information on configuration and measurement control for different sensors.
Chapter 1 _________________________________________________________ General Information Feedback Vaisala Customer Documentation Team welcomes your comments and suggestions on the quality and usefulness of this publication. If you find errors or have other suggestions for improvement, please indicate the chapter, section, and page number. You can send comments to us by e-mail: manuals@vaisala.
Technical Reference_________________________________________________________________ Trademarks Microsoft, Windows, and Windows NT are registered trademarks of Microsoft Corporation in the United States and/or other countries. License Agreement All rights to any software are held by Vaisala or third parties. The customer is allowed to use the software only to the extent that is provided by the applicable supply contract or Software License Agreement.
Chapter 2 ________________________________________________ Configuring Optional Hardware CHAPTER 2 CONFIGURING OPTIONAL HARDWARE This chapter contains information needed when configuring optional hardware. The optional hardware described here includes Vaisala Digital I/O Module and Nokeval 7470 Serial to Analog Converter. Digital I/O Module Digital I/O Module QMI108 is used to extend the I/O capacity of MAWS. It is equipped with eight inputs and eight outputs.
Technical Reference_________________________________________________________________ 0506-036 Figure 1 2. Hardware Setup View: Digital I/O Configuration Proceed to the Timers view to configure the on/off cycle for the selected outputs (0 and 4).
Chapter 2 ________________________________________________ Configuring Optional Hardware In the example, the configuration is as follows: - Test output 0 toggles between on and off in a one-minute cycle. - Test output 4 is on for one minute before each full hour. Alternatively, you can control the outputs with alarms, for example, by activating the output signal when the alarm condition is met. Proceed as follows: 1.
Technical Reference_________________________________________________________________ In the Reports view, the input statuses are available as individual variables IN0 through IN7 and combined as bits into a single status word ALL. The input module status is also available as for the sensors, refer to Figure 4 below.
Chapter 2 ________________________________________________ Configuring Optional Hardware current inputs are used instead of a serial data interface. The converter can be controlled through an RS-485 serial line. As Nokeval converters are fully configurable, knowledge on these devices is required before taking them into use with MAWS. It is possible to add one or more converters to a single RS-485 serial line with MAWS. For further details, see the applicable Nokeval user's manual, available at www.nokeval.
Technical Reference_________________________________________________________________ 0412-069 Figure 6 4.
Chapter 2 ________________________________________________ Configuring Optional Hardware 0412-070 Figure 7 Hardware View: Configuring Communication Port for Nokeval 7470 (2/2) 5. In the Buffers frame, leave the default values. 6.
Technical Reference_________________________________________________________________ 7. NOTE Select Nokeval Converter from the Aux list and connect it to the selected serial port. Add only one Nokeval Converter to the serial port here, even if there are several devices connected to this serial line. Do not connect more than one Nokeval Converter to one serial port in the Equipment view.
Chapter 2 ________________________________________________ Configuring Optional Hardware Give the address as an 8-bit hexadecimal value where the highest bit is always set. The hexadecimal value is given in the report using the special character code \x. For example, if the device address is 0, then the first report field should start with \x80. If the device address is 2, then the field starts with \x82. 3. Give the Nokeval command, excluding the data value, directly after the address.
Technical Reference_________________________________________________________________ 0506-042 Figure 10 Communications View: Connecting the Report to Nokeval 7470 Scaling of Data All data values must be properly scaled to fit into the output range of the Nokeval converter outputs. You can carry out the scaling using two alternative methods: - Configuring MAWS to scale the data. This is performed using a User definable conversion, configured in the Calculations view of MAWS Lizard.
Chapter 2 ________________________________________________ Configuring Optional Hardware Alternatively, any specific data reading can be sent when the data value is invalid, by specifying an invalid data indicator for the data cell in the report. Timing of Reports Up to 32 reports can be generated successively at the same synchronization time. MAWS will queue the commands and process them one by one.
Technical Reference_________________________________________________________________ Table 3 Nokeval 7470 Specifications Parameter Current output: Outputs Maximum load Voltage output: Outputs Maximum load Accuracy Linearity Temperature effect Response time of output RS-485/RS-232 input DAC conversion Power supply Power consumption Power demand IP protection Mounting Weight Value/Description Common negative wire 0 ... 20 mA, 4 ... 20 mA 600 Ω Common negative wire 0 ... 5 V, 0. ... 10V 1 mA 0.
Chapter 3 ___________________________________________________________ Managing Setups CHAPTER 3 MANAGING SETUPS This chapter provides additional information on setup management. Setup Memory Size For software version 5.0.1 or higher, the size of the setup memory is 224 kilobytes, allowing you to create also larger setups. The internal logging memory size is 1 638 400. Executing a Setup File from Compact Flash Memory Card With the MAWS software version 5.
Technical Reference_________________________________________________________________ HydroMet HelpDesk. You will need to send your logger to Vaisala for the update. Currently there are external memory cards available from 32 megabytes (MB) up to hundreds of megabytes. Memory cards over 64 megabytes are not recommended by Vaisala. Cards can be read directly in the PC.
Chapter 3 ___________________________________________________________ Managing Setups 3. To check that a configuration exists, give the following command in MAWS Terminal. Note that MAWS uses a UNIX/LINUX type file system with only one root and thus automatically mounts the CF card on the /Ext/ directory path. dir /Ext/ where pathname 4.
Technical Reference_________________________________________________________________ . rwxd 13:11:40 21.03.2005 0 .. rwxd 13:11:40 21.03.2005 0 setup501.adc rw-- 12:26:28 21.03.2005 Files shown: 3 1776 Bytes used: 1776 / W> reset 0 /Ext/TEST/setup501.adc Resetting device in 0 seconds COM0: QML201 Startup - Warm Boot Serial # : Z402074 Hardware : Rev B-001 Software : 5.01 Checksum : 00000000 Boot sw version : 4.07 System RAM : 2048kB Free memory : 1768kB Internal temp. : 23.
Chapter 3 ___________________________________________________________ Managing Setups Executing an Alternative Setup File from Compact Flash Memory Card You can execute an alternative setup file from the Compact Flash (CF) memory card. This is especially useful when upgrading new configurations remotely. An alternative setup file will be executed if the primary setup file in the /Cfg directory fails. NOTE You must set a specific static parameter to enable this.
Technical Reference_________________________________________________________________ To verify that the path is correct, give the SPSET command. An output from MAWS is as follows: / > SPSET alternativescript = /Ext/test/setup501.adc If an error occurs while executing the primary setup file, MAWS will check whether the static parameter alternativescript is set. If so, MAWS will be reset. After the reset, MAWS will check whether an alternative setup file exists.
Chapter 3 ___________________________________________________________ Managing Setups COM0: QML201 Startup - Cold Boot Serial # : Z402074 Hardware : Rev B-001 Software : 5.01 Checksum : 00000000 Boot sw version : 4.07 System RAM : 2048kB Free memory : 1768kB Internal temp. : 23.23'C Active errors : NO Active warnings : NO Piggyback - 0 : DSU232 rev: B serial no: 008232 Piggyback - 1 : DSU232 rev: B serial no: 003523 NEW flash checksum (02400000 to 025FFFFF) is BBCC1130 Found setup /Cfg/failing.
Technical Reference_________________________________________________________________ This page intentionally left blank.
Chapter 4 ____________________________________________ Configuring Communication options CHAPTER 4 CONFIGURING COMMUNICATION OPTIONS Iridium Satellite Transmitter This chapter provides information on the Iridium satellite transmitter communications option, how to interface the Iridium satellite transmitter to MAWS, and more specifically, how to use the Short Burst Mode in binary transmission. Iridium transmission using data calls can be used like any other modem.
Technical Reference_________________________________________________________________ Configuring Iridium Transmission Iridium transmission is configured using MAWS Lizard. Proceed as follows: 1. Start MAWS Lizard. 2. Open an existing setup in the Setup view or create a new one. 3.
Chapter 4 ____________________________________________ Configuring Communication options 4. Proceed to the Equipment view by clicking Next. 5. Click Aux to list the communication devices, see Figure 12 below. - Scroll down the Available list for Iridium. - Click the Add button and select Iridium. - Double-click the Connector list for Iridium_modem_1 in the I/O Connections frame in order to select the COM port. - Leave the EXCITATION selection to .
Technical Reference_________________________________________________________________ c. Click the Link button to make a connection between the report and the Iridium modem. d. Modify the parameters in the Configure frame on the right. 0509-053 Figure 13 Communications View: Configuring Iridium Transmission Parameters In order to adjust the parameters, refer to Configuring Modem Options Technical Reference. As a default, the parameters are set for using Short Burst Data (SBD).
Chapter 4 ____________________________________________ Configuring Communication options 0509-054 Figure 14 Communications View: Configuring Iridium Transmission Parameters Using Power Control Option The Powering device option in the Options frame lists all the Power Control components in the setup. For more detailed information on how to use this option, see section Power Control Option on page 41. 7.
Technical Reference_________________________________________________________________ 0509-055 Figure 15 Communications View: Selecting Report Transmission Options for Iridium For the other options, make the same selections as for any PSTN modem. The fields Send to (on success and on fail) are currently used only when transmitting via SBD. You should leave the fields empty unless otherwise informed. Iridium Configuration Options There are also configuration options for Iridium.
Chapter 4 ____________________________________________ Configuring Communication options 0509-056 Figure 16 Communications View: Selecting Iridium Configuration Options VAISALA________________________________________________________________________ 35
Technical Reference_________________________________________________________________ This page intentionally left blank.
Chapter 5 ___________________________________________ Configuring Calculations and Reports CHAPTER 5 CONFIGURING CALCULATIONS AND REPORTS Accumulator with Predefined Reset Time You can use the "Accumulator" component to add a source parameter to the destination parameter in order, for example, to accumulate annual precipitation in the MAWS system and to reset the calculation at a predefined moment.
Technical Reference_________________________________________________________________ 0511-113 Figure 17 2. Calculations View: Selecting the Accumulator Component Select the Accumulator in the list by clicking on the item name and then the Add button. Configuring Accumulator To configure the Accumulator, proceed as follows: 1.
Chapter 5 _________________________________________________________________________ 3. You can also choose a reset option (setting of the static parameter to zero at a selected time) in the Reset time frame. If you select the No reset check box, the destination parameter is incremented indefinitely unless you reset the Accumulator manually. If No reset is unselected, click on the Reset time button to select the resetting schedule.
Technical Reference_________________________________________________________________ Resetting Accumulator Manually You can also reset the Accumulator manually, that is, assign a positive initial value of zero or greater than zero to the accumulation. In order to reset the Accumulator when using MAWS Terminal, proceed as follows: 1. Start MAWS Terminal and select the Station Settings option from the Tools menu or click on the Set Station Settings icon in the Toolbar, see Figure 19 below. 2.
Chapter 5 _________________________________________________________________________ where static parameter = initial value = Name of the static parameter, selected from the Select static parameter combo box or entered manually in the text box in the Configure frame of the Calculations view. The initial value you wish to assign for the accumulation. Always enter the value with one decimal place (for example, 0.0).
Technical Reference_________________________________________________________________ Configuring Power Control Option To configure the Power Control option, proceed as follows: 1. Start MAWS Lizard. 2. Open an existing setup in the Setup view or create a new one. 3. Proceed to the Calculations view to configure Power Control, see Figure 20 below. 4. Click Misc to list the available options. 5. Select PowerCTRL from the Available list and click Add. 8.
Chapter 5 _________________________________________________________________________ Timer Option Timer is used to create timed events that occur at a predefined time at specific intervals. These events resemble the timers in the Timers view but are more freely configurable to meet specific needs. The Timer option is located in the Misc list in the Select Calculation frame. For a variety of options, see also the Conv list. Configuring Timer Option To configure the Timer option, proceed as follows: 1.
Technical Reference_________________________________________________________________ There are several configuration options in the Configuration frame on the right: Option Repeat times Scale Time Zone Priority Sync time [ms] Trig interval time [ms] Description Defines how many timed events will be created Indicates whether the timed events are tied to absolute time or whether they are relative and only the interval is meaningful Specifies whether UTC or local time is used to calculate the timed events,
Chapter 5 _________________________________________________________________________ 0509-059 Figure 22 Reports View: UNIX Time Option under Time Snap Checksum Option in Reports With the MAWS software version 5.02 or higher, it is possible to add a checksum into the report that is already in the report configuration. You can also define where the data begins and ends in the checksum calculation as well as format the checksum fields to meet your own requirements.
Technical Reference_________________________________________________________________ 0509-060 Figure 23 Reports View: Configuring New Checksum Report There are new variables in the list that are used for calculating and showing the checksum: - CRC returns the checksum to be formatted in a User format - CRC_start indicates the start of data to be included in the CRC - CRC_end stops calculating the CRC - CRC_string returns the checksum in an ASCII presentation 46 __________________________________
Chapter 5 _________________________________________________________________________ To calculate a CRC, proceed as follows. Note that this procedure is an example of a typical case. 1. Go to the Reports view, see Figure 24 below: 0509-061 Figure 24 Reports View: Adding CRC The variables from rows 2 to 6 are taken into account when calculating the CRC. The CRC is placed after the ******* string. 2. For the CRC calculation method, see the Report properties frame.
Technical Reference_________________________________________________________________ 0509-062 Figure 25 Checksum Settings Window In the Checksum settings window you can define the following parameters: - Method Parameter Option SUM XOR CRC16 CRC32 USER CRC Description Calculates a sum from the bytes to be included in CRC Calculates an exclusive OR from the bytes to be included in CRC 16-bit CRC 32-bit CRC User-defined CRC - Format Parameter Option Hex Hex LSB first KERMIT WXT Description Displays the
Chapter 5 _________________________________________________________________________ - Used bitcount, specifies how many bits are used to calculate the CRC. 16 bits is typical for SUM and CRC16, 8 bits is typical for XOR. - Init value, indicates the start value for the CRC, that is, where the calculation is started. This parameter is also known as a seed.
Technical Reference_________________________________________________________________ 0509-063 Figure 26 Reports View: Selecting Binary Report Form A binary report is configured like any other report. Binary reports differ from regular ones in the formatting of the data items. Each time a new item is added to a report, a formatting frame appears, see Figure 27 on page 51.
Chapter 5 _________________________________________________________________________ 0509-064 Figure 27 Reports View: Binary Report Formatting Frame (1/2) Scaling parameters define how the data item will be scaled into the binary report: - Unit conversion, specifies whether the value will be converted to other units, for example from Celsius degrees to Kelvin. This parameter functions in the same way as in a regular report.
Technical Reference_________________________________________________________________ Parameter Option Integer Float Double BCD - Description Presents the value both as a negative and a positive integer value Presents the value as a floating point number using 4 bytes: sign bit, 8-bit exponent, 23-bit mantissa Presents the value as a floating point number using 8 bytes: sign bit, 11-bit exponent, 52-bit mantissa Presents the value as a binary code decimal Invalid data, used to indicate which value will b
Chapter 5 _________________________________________________________________________ - Include parity bit, indicates that a parity bit is added to this field to assure data validy. This option increases field length by one.
Technical Reference_________________________________________________________________ - Bit count, indicates how many bits are included in this report - Byte count, indicates how many bytes are included in this report - Zero, return number zero. This parameter can be used to add a constant into the binary report.
Chapter 6 _________________________________________________________ Configuring sensors CHAPTER 6 CONFIGURING SENSORS This chapter provides advanced information on configuration and measurement control for different sensors. For basic information on MAWS setups, refer to MAWS Lizard Setup Software User's Guide.
Technical Reference_________________________________________________________________ WXT510 to be used with MAWS has the following default configuration: Checksum Automatic precipitation reset Mode Precipitation message ON ON Autosend Time-based For more detailed information on the measured parameters, see Weather Transmitter WXT510 User´s Guide.
Chapter 6 _________________________________________________________________________ However, if you need to check/modify the configuration, proceed as follows. WXT510 is configured using MAWS Lizard. 1. Start MAWS Lizard. For example, click on the Start button and select Programs, Vaisala and MAWS Lizard. 2. Open an existing setup in the Setup view or create a new one. 3. Proceed to the Hardware view by clicking Next to configure the WXT510 sensor hardware, see Figure 29 below.
Technical Reference_________________________________________________________________ 4. Proceed to the Equipment view by clicking Next, see Figure 30 below. - Scroll down the Available list box for WXT510. - Click on the Add button to select WXT510. - Double-click the Connector list for WXT510 in the I/O Connections frame in order to select the COM port. 0504-029 Figure 30 5.
Chapter 6 _________________________________________________________________________ 0504-030 Figure 31 Measurements View: Configuring WXT510 Communications Options There are several parameter options you can set in the Communications frame of the Measurements view. Note that in general, the predefined settings are the most suitable for a typical MAWS setup. In order to change a parameter option, you simply need to select or unselect the corresponding check box.
Technical Reference_________________________________________________________________ Select also the Messages in use: - Wind: wind speed and direction are received - Supervisor: internal data is received from WXT510 - Precipitation: the amount, duration and intensity of rain and hail are received - PTU: air pressure, temperature and humidity are received Note that the options you select must be enabled in the connected WXT510.
Chapter 6 _________________________________________________________________________ With the communication timeout parameters in Cyclic transmit mode, you can define a maximum interval between two messages. If the new message from WXT510 has not been received within the given time, the data in that message is then marked as invalid. Using Sensor Variables in Reports You can use the variables received from the Weather Transmitter WXT510 in MAWS Lizard like those from any other sensor.
Technical Reference_________________________________________________________________ 0504-032 Figure 33 Reports View: Checking WXT510 Variables Gill's WindSonic To configure Gill's WindSonic, proceed as follows: In the Equipment view, select WindSonic from the sensor list and connect it to the selected serial port. After that, the sensor is ready for use. NOTE The WindSonic transmits data continuously, 4 times per second.
Chapter 6 _________________________________________________________________________ There are several serial output settings available for this sensor.
Technical Reference_________________________________________________________________ Configuring Ultrasonic Anemometer USA-1 Communications Setup The USA-1 wind sensor is connected to MAWS through a RS-422 or RS-232 serial line. Follow METEK's instructions when selecting cabling for the desired mode. The default data transmission parameters for USA-1 are the following: 9600 bps, 8 data bits, no parity, 1 stop bit.
Chapter 6 _________________________________________________________________________ 0504-033 Figure 34 4. Hardware View: Configuring USA-1 Scroll down the Configure frame until the following options are displayed, see Figure 35 below.
Technical Reference_________________________________________________________________ 5. Select the Ignore frames check box in the Misc frame. This option disables handling of messages received from USA-1 during startup and thus prevents receive overflow. 6. Proceed to the Equipment view by clicking Next, see Figure 36 below. - Scroll down the Available list box for USA-1. - Click on the Add button to select USA-1.
Chapter 6 _________________________________________________________________________ 3. Click Metek_1 in the Available list box in the Select Variables frame. 4. Lizard displays the items that are available for the source in question. The items are grouped under source names like files in directories. To see the list of items under a source, click on the plus (+) sign. When you click on an item in the Select Variables list box, a brief description also appears in the information pane on the lower left.
Technical Reference_________________________________________________________________ 2. Status, indicates the internal status of the MAWS interface for USA-1, and the possible values.
Chapter 6 _________________________________________________________________________ 0412-027 Figure 38 Wiring Diagram for QMI108 with WAA151 and WAV151 - Sensors Powered Continuously 0412-026 Figure 39 Wiring Diagram for QMI108 with WAA151 Only VAISALA________________________________________________________________________ 69
Technical Reference_________________________________________________________________ 0412-024 Figure 40 Wiring Diagram for QMI108 with WAV151 Only The wind vane is added to the setup in the Equipment view. Simply select WAV151 from the sensor list. There is no need for other connections because WAV151 is always connected to the QMI108 unit. Typically, WAV151 is continuously powered from an external 12 VDC power source. In this default mode, there is no need to set or change any default parameters.
Chapter 6 _________________________________________________________________________ 0412-067 Figure 41 Measurements View: Configuring Powering Control For the wiring diagram, see Figure 42 below.
Technical Reference_________________________________________________________________ - In the Powering polarity frame: The QMI108 output, which is typically normal but can be reversed. This depends on the hardware used to control powering. - The Pre-measure delay field: The delay between power on and measurement start [ms]. This option provides stabilization time for sensor before measurement is made. The value should never be less than 60 milliseconds.
Chapter 6 _________________________________________________________________________ HMT330 is configured using MAWS Lizard Setup Software. Proceed as follows: 1. Start MAWS Lizard. 2. Open an existing setup in the Setup view or create a new one. 3. Proceed to the Hardware view by clicking Next to configure serial transmission hardware, see Figure 43 below. In this example, MOD1/1 on DSU232 is selected for use with the HMT330 sensor.
Technical Reference_________________________________________________________________ 0509-048 Figure 44 5.
Chapter 6 _________________________________________________________________________ There are several parameter options you can set in the Communications frame of the Measurements view. Note that, in general, the predefined settings are the most suitable for a typical MAWS setup. In order to change a parameter option, you simply need to select or unselect the corresponding check box. The following options are available: - Use the Device ID (editable) text box to distinguish HMT33.
Technical Reference_________________________________________________________________ 4. Lizard displays the items that are available for the source in question. For more information on these variables, refer to Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330 User's Guide. 0509-050 Figure 46 Reports View: Selecting HMT330 Variables Temperature Measurement PT1000 With MAWS software, you can measure PT1000 elements.
Chapter 6 _________________________________________________________________________ 0412-066 Figure 47 Equipment View: Selecting I/O Connections for PT1000 By default, the PT1000 measurement is a 3-wire connection but you can change it to a 4-wire connection on the Advanced user level by changing the measurement type. Typically, you do not need to change the predefined PT1000 sensor parameters. Only the data validation parameters can vary, depending on the installation site.
Technical Reference_________________________________________________________________ 0412-022 Figure 49 Wiring Diagram for 4-Wire Connection Temperature Measurement PT100 in 3-Wire Connection The PT100 sensor in a 3-wire connection is rarely used but can be configured with MAWS Lizard like any other sensor. In the Equipment view, select the predefined PT100 sensor from the list. Proceed then to the Measurements view and select a PT100 measurement.
Chapter 6 _________________________________________________________________________ See also Configuring QSE101 Shaft Encoder with MAWS Lizard Technical Reference. Figure 50 Absolute Shaft Encoder QSE104 Configuring Absolute Shaft Encoder QSE104 To add the QSE104 sensor into your system setup, proceed as follows: 1. Go to the Equipment view of MAWS Lizard and select QSE104 from the sensor list, see Figure 51 on page 80. 2.
Technical Reference_________________________________________________________________ 0412-046 Figure 51 3. Equipment View: QSE104 Connected to the Optional DSI486 Module Click Next or go to the Measurements view. All the parameters are predefined. If there are more than one device on the same RS-485 line, the device IDs must be different. See Figure 52 below.
Chapter 6 _________________________________________________________________________ Setting Current Water Level for Absolute Shaft Encoder You can read or set the current water level, that is, the current data value of the shaft encoder, using the following command in MAWS Terminal: waterlevel [new value] If you enter the parameter new value, the current data value is set as the parameter value. If you do not enter the parameter, then the current data value is displayed.
Technical Reference_________________________________________________________________ 0506-044 Figure 53 MAWS Lizard Waterlevel Window For example, to set the current waterlevel to 6.120 meters for the QSE104_2 sensor, give the following command in MAWS Terminal: /> waterlevel_2 6.120 water level = 6.1200 m Geonor Water Gauge The Geonor T-200B3 precipitation gauge contains three separate vibrating wire sensors.
Chapter 6 _________________________________________________________________________ 0412-029 Figure 54 Example Wiring Diagram for Geonor T200 B3 VAISALA________________________________________________________________________ 83
Technical Reference_________________________________________________________________ The sensor has three calibration coefficients that are supplied by the manufacturer of the sensor. They must be given to MAWS using the following static parameters: 1. Three calibration coefficients for each sensor, see Table 4 below. Table 4 Sensor Sensor 1 Sensor 2 Sensor 3 2.
Chapter 6 _________________________________________________________________________ These parameters can be set in one of the following three ways: a. In Lizard Setup view, in the Static Parameters tab.
Technical Reference_________________________________________________________________ 0412-049 Figure 56 b. Measurements View: Configuring Geonor Using MAWS or PALM Terminal. There is a static parameter editor in the Station Settings window. See Figure 57 below.
Chapter 6 _________________________________________________________________________ c. Using maintenance connection and the spset command. The following example sets parameters for sensor 1: >\ spset geonort200_f0_1 1051.4 >\ spset geonort200_A_1 0.016920 >\ spset geonort200_B_1 0.000009226750 Configuring Geonor Water Gauge T-200B3 To configure the Geonor T-200B3 sensor, proceed as follows: 1. Go to the Equipment view in MAWS Lizard. 2.
Technical Reference_________________________________________________________________ In the Algorithm frame, the first option Averaging is a Vaisala proprietary algorithm. The USCRN precipitation algorithm is a US Climate Reference Network algorithm that is published on the USCRN www pages. Valid sensors box specifies how many of the three sensor measurements must have VALID status before the final calculation of the results.
Chapter 6 _________________________________________________________________________ Table 6 Output Parameters Parameter ID Type PR Double PRsum Double evaporation Double fullness Double freq_1 freq_2 freq_3 PRsum_1 Double Double Double Double PRsum_2 Double PRsum_3 Double PR_1 Double PR_2 Double PR_3 Double evaporation_1 Double evaporation_2 Double evaporation_3 Double Description Precipitation detected and reported by selected algorithm in user-set calculation interval Precipitation a
Technical Reference_________________________________________________________________ 0511-028 Figure 60 Timers View: Adjusting Timers for Geonor T-200 Rim Heating The rim heating software component is designed to be used with the Geonor T-200B3 precipitation gauge. This component offers several parameters to control the heating element of the Geonor T-200B3 sensor. There are no restrictions for using this component elsewhere if similar type of control is required.
Chapter 6 _________________________________________________________________________ 0412-021 Figure 61 Control Flow for Heating Selecting and Configuring HeatControl Component The HeatControl component is available in the Calculations view, see Figure 62 on page 92. 1. In the Select Calculation frame, click the Misc button to display a list of miscellaneous calculations. 2. Select HeatControl from the list and click Add. 3. HeatControl_x appears in the Selected frame.
Technical Reference_________________________________________________________________ 0511-110 Figure 62 4. Calculations View: Selecting HeatControl To configure the HeatControl component, click on HeatControl_x in the Selected list.
Chapter 6 _________________________________________________________________________ The first part of the configuration includes the mandatory parameters and some advanced parameters needed only if an Additional check is selected. See Figure 64 on page 93.
Technical Reference_________________________________________________________________ - Heater channel to select the control output for the heating element. Typically, the QMI102 digital I/O unit is used to control the heating element, but it is also possible to use the excitation pins of the channels CH0..3 and CH A/B . The HeatControl component is activated by a user-definable timer, configured like any other timer.
Chapter 6 _________________________________________________________________________ Lambrecht QUATRO-IND Lambrecht QUATRO-IND is a weather instrument that measures the following variables: - Wind speed and direction - Air temperature - Humidity and dewpoint - Air pressure Configuring Lambrecht QUATROIND Communications Setup The QUATRO-IND weather sensor is connected to MAWS through a RS-485 serial line.
Technical Reference_________________________________________________________________ 3. Go to the Hardware view, see Figure 66 below. 0412-058 Figure 66 Hardware View: Configuring QUATRO-IND (1/2) In the example above, COM1 has been selected for use with the QUATRO-IND sensor. Under Line parameters the settings are: 4800, 8, 1, none. NOTE Scroll down for Transmit Control to select Transmit control enabled, as otherwise MAWS keeps the tx driver continuously ON and receiving is not possible.
Chapter 6 _________________________________________________________________________ 0412-059 Figure 67 4. Hardware View: Configuring QUATRO-IND (2/2) Select Next to go to the Equipment view, see Figure 68 on page 98. - Scroll down the Available list to find Lambrecht QUATROIND. - Click the Add button to select Lambrecht QUATRO-IND. - Double click the Connector field for Lambrecht in order to select the COM port.
Technical Reference_________________________________________________________________ 0412-060 Figure 68 5. Equipment View: Selecting I/O Connection Select Next to go to the Measurements view, see Figure 69 below. 0412-061 Figure 69 Measurements View: Configuring QUATROIND There are four internal pieces of MAWS equipment for QUATROIND, one component for each measurement to be reported, see Figure 69 above.
Chapter 6 _________________________________________________________________________ measurement, the information pane shows the actual sensor. You can separately enter the Timeout value for each measurement; the default is 3 seconds. Next, you can use the measurements of QUATRO-IND as usual.
Technical Reference_________________________________________________________________ Sensor Monitoring To test if the communication between MAWS and the QUATRO-IND sensor works properly, proceed as follows: 1. Start MAWS Terminal and open the maintenance connection. 2. Check communication by opening a connection between the maintenance port and sensor. When supposing that the QUATRO-IND sensor is connected to COM1).
Chapter 6 _________________________________________________________________________ Each parameter is updated once a second, so you should see four messages received every second. If no data is received, check that: - QUATRO-IND is powered up, - wires are correctly connected, and - serial parameters are correctly set in MAWS Lizard. Finally, close the terminal connection: /> close 3. Check that MAWS receives parameters from QUATRO-IND correctly.
Technical Reference_________________________________________________________________ In case of an error, the response can be one of the those listed in Table 11 below. Table 11 Error Responses and Probable Causes Response Target component not found Invalid ID Status:2 Value:.. Probable Cause The component is misspelled or QUATRO-IND has not been configured The data item name is misspelled or the F flag at the end of the command is missing The parameter is not received correctly.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces CHAPTER 7 CONFIGURING SERIAL SENSOR INTERFACES This chapter contains information needed when configuring serial sensor interfaces, that is, connecting sensors that use serial communications to the MAWS301 weather station. Note that this chapter contains only instructions for configuring the interface, not the sensor itself.
Technical Reference_________________________________________________________________ For most of the sensors, wiring depends on the application. You should therefore also have additional wiring instructions and drawings available. Physical Interface MAWS logger has five different types of optional serial interfaces available for the sensor connections. For details and wiring for the interface modules, refer to the MAWS user manuals listed in the Related Manuals table on page 6.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces Data Acquisition Methods In general, there are two alternate methods to connect a serial sensor to the MAWS weather station: 1. Automatic transmission - Device transmits the measurement results automatically - Only one device per serial port - Physical connection can be RS-232 or RS-485 - Higher data throughput 1.
Technical Reference_________________________________________________________________ Configuring Serial Line Parameters NOTE MAWS does not perform any automatic configuration of the connected sensors. You have to perform the configuration and store it manually by using the sensor's own service interface. For sensors without a human readable command interface, extra MAWS shell commands are provided. For further information, refer to detailed sensor interface descriptions or separate Technical References.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces Port DSU232 / COM1/1 DSU232 / COM1/2 DSU232 / COM2/1 DSU232 / COM2/2 DSI485 / COM1 DSI485 / COM2 DSI486 / COM1/1 DSI486 / COM1/2 DSI486 / COM2/1 DSI486 / COM2/2 Name DSU232_0_0 DSU232_0_1 DSU232_1_0 DSU232_1_1 DSI485_0 DSI485_1 DSI486_0_0 DSI486_0_1 DSI486_1_0 DSI486_1_1 The following example describes how to access the PWD11 sensor with device identifier 1, connected to the COM2 / 2 port using a DSI486 piggyback.
Technical Reference_________________________________________________________________ Configuring Sensor Interface Serial sensors have and use their own configuration interfaces. In addition to defining the communication parameters, there is typically very little to configure with Lizard setup software. Proceed as follows: 1. Add and configure the communication interface module using the Hardware Setup view, see Figure 71 below.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces 0506-051 Figure 72 Hardware Setup View: Defining Transmit Control Parameters You do not need to select the Service connection enabled option in order to have service access to the sensor. Selecting this option in fact enables service access to MAWS through the port. 2. Proceed to the Equipment view to add and connect the sensor, see Figure 73 on page 110.
Technical Reference_________________________________________________________________ 0506-052 Figure 73 Equipment View: Adding and Connecting the Sensor Note that serial sensors do not have default connectors, so you need to connect the sensor manually to the selected port. The first number of the connector identifier MODX/X (for example, MOD1/1) indicates the piggyback slot: 1 is the one closer to the battery holder and 2 is the bottom one.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces 0506-053 Figure 75 Measurements View: Configuring the Sensor Interface Typical items to configure include the following: Item Device ID Mode Message Type Timeout NOTE Description / Use Device identifier (must be entered) - Same as the set - Unique for each sensor if multiple devices are connected to the same serial line Used by - Polling and data reception to identify the device - Open commands to identify whi
Technical Reference_________________________________________________________________ Serial Sensor Interfaces This chapter describes the following options in detail for each sensor: - Default communication parameters - Typical sensor interface configuration alternatives - Data availability with different message types. This refers to the availability of sensor output variables used in the setup for calculation, logging and reporting.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces Available data with FD12P: Lizard variable Message Description pwd_id dev_id alarms vis_1 vis_10 pw_nws pw_inst pw_15 pw_1h water_1h water_cum snow_cum t_xarm lumin metar_i metar_r pw_insttxt 2,7 2,7 2,7 2,7 2,7 2,7 2,7 2,7 2,7 2,7 2,7 2,7 7 7 7 7 2,7 pw_15txt 2,7 pw_1htxt 2,7 water_inc 2,7 snow_inc 2,7 orig_PWD 2,7 Header text Device ID Device alarm code Visibility 1 min average Visibility 10 min aver
Technical Reference_________________________________________________________________ NOTE Lizard variable Message Description water_cum snow_cum t_xarm lumin metar_i metar_r pw_insttxt 2,7 2,7 7 n/a 7 7 2,7 pw_15txt 2,7 pw_1htxt 2,7 water_inc 2,7 snow_inc 2,7 orig_PWD 2,7 Cumulative water sum Cumulative snow sum Crossarm temperature Background luminance Instant METAR weather code Recent METAR weather code WMO text for instant present weather code WMO text for 15 min present weather code WMO
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces Sensor commands for interface configuration: Operation Set polled Command(s) SET MESSAGE TYPE SET MESSAGE MODE POLLING Set auto tx SET MESSAGE TYPE SET MESSAGE MODE AUTOSEND Set ID SET UNIT_ID Set interface type SET PORT DATA INTERFACE Purpose / Note Set message type, where is MSG1, MSG6 or MSG 61. Disables automatic transmission.
Technical Reference_________________________________________________________________ Lizard Variable Message Description oktas_3 height_3 oktas_4 height_4 oktas_5 height_5 d_stat base_1 base_2 base_3 v_vis h_sig i_stat orig_CT 6,61 6,61 6,61 6,61 61 61 1,6,61 1,6,61 1,6,61 1,6,61 1,6,61 1,6,61 1,6,61 1,6,61 Octas in layer 3 Height of layer 3 Octas in layer 4 Height of layer 4 Octas in layer 5 Height of layer 5 Detection status Height of cloud base 1 (* Height of cloud base 2 (* Height of cloud base 3
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces Operation Set ID Command(s) ADDR Activate param RESET Purpose / Note Set device id for polled operation. = 0 ... 99 New parameters are taken into use after reset. To get a suitably formatted data output from the PTB220 sensor, the following message format string must be given using the FORM command. Note that the string is almost of maximum length allowed, so do not insert any extra spaces.
Technical Reference_________________________________________________________________ Ultrasonic Wind Sensor WS425 Default communication parameters: Parameter Speed Parity Data bits Stop bits Handshake Value 9600 bps none 8 1 none The sensor has a menu-driven configuration interface, which is shown right after the service connection has been opened with the open command.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces WT500 Series Wind Transmitters The WT500 series wind transmitter can be used to interface with different types of wind sensors, and optionally with the humidity probe HMP45 for temperature and humidity measurement.
Technical Reference_________________________________________________________________ NOTE MWVQUERY format can only be used alone. If you are using XDR_TU, you must use the MWV format for wind data transfer, and operate WT5XX in an automatic transmission mode. NOTE After configuring the device, enter the RESET command in order for the changes to take effect.
Chapter 7 ____________________________________________ Configuring Serial Sensor Interfaces GARMIN GPS35-PC GPS Receiver Default communication parameters according to NMEA 0183: Parameter Speed Parity Data bits Stop bits Handshake Value 4800 bps none 8 1 none No initial configuration is needed. The device automatically starts outputting GPRMC data at one-second intervals.
Technical Reference_________________________________________________________________ This page intentionally left blank.
Chapter 8 ___________________________________________________________________ Alarms CHAPTER 8 ALARMS This chapter provides information on output control using alarms. Output Control Using Alarm A free measurement channel excitation output from channels CH0 to CH3 can be used as a digital output, although MAWS has no real digital outputs. The Set excitation output alarm provides a means for controlling these outputs using the MAWS alarm functionality.
Technical Reference_________________________________________________________________ 0208-054 Figure 76 2. Configuring the Excitation Output Control Select the Alarm view and create a new alarm with the type Set excitation output, refer to Figure 77 below. 0208-055 Figure 77 3.
Chapter 8 ___________________________________________________________________ Alarms - You do not need to enter any separate alarm on and off targets. - Output is always on as long as the alarm condition exists, that is, the option for triggering the action on each test is missing.
Technical Reference_________________________________________________________________ This page intentionally left blank.
Chapter 9 __________________________________________________________ Additional features CHAPTER 9 ADDITIONAL FEATURES This chapter provides information on other features or enhancements to MAWS Lizard. Floating Point Numbers in Static Parameters In MAWS version 4.07 or earlier, floating point numbers are stored in the static parameters as 32-bit floating point numbers (float). In MAWS version 5.01 or higher, the floating point numbers use a 64bit floating-point number (double).
Technical Reference_________________________________________________________________ 0412-068 Figure 78 Measurements View: Selecting the Frequency Mode Passive FTP Mode with iConnector Due to firewall configurations, it is sometimes necessary to use a passive FTP mode to send files to an FTP server. When you select the Use passive mode FTP option, iConnector uses passive mode FTP sessions. Hence, iConnector uses the PASV command instead of PORT when opening an FTP session.
Chapter 9 __________________________________________________________ Additional features 0412-075 Figure 79 Communications View: Use Passive Mode FTP Option VAISALA_______________________________________________________________________ 129
Technical Reference_________________________________________________________________ This page intentionally left blank.
Appendix A _________________________________________________________ Sensor Status List APPENDIX A SENSOR STATUS LIST Table 13 Value 0 1 2 3 4 5 6 7 8 ... 19 20 21 22 23 24 25 26 27 28 29 30 99 Sensor Status List Meaning The sensor is working properly (OK) Not measured yet Interface is not initialized Communication timeout has occurred Unknown data is received Communication is functioning, but the sensor reports errors.
Technical Reference_________________________________________________________________ 1) Value is available only for the sensors with the serial interface. 2) Value is available only for the sensors with a conventional, that is, analog or counter/frequency interface.
_____________________________________________________________________________Index INDEX A Absolute Shaft Encoder adding this sensor to your system setup 79 introduction 78 setting current water level 81 Accumulator with predefined reset time configuring 38 introduction 37 resetting manually 40 using in reports 41 Additional features Enhanced generic frequency measurement 127 Floating point numbers in static parameters 127 Using passive FTP mode with iConnector 127 Alarms 123 output control using alarms 1
Technical Reference_________________________________________________________________ E License agreement Enhanced generic frequency measurement127 Enhanced Time Snapshot introduction 44 Executing a setup file from CF memory card introduction 21 Executing a setup from CF memory card Quick Reference 24 testing a setup 22 Executing an alternative setup file from CF memory card enabling this feature 25 introduction 25 M F Feedback 7 Floating point numbers in static parameters127 G GARMIN GPS35 available da
_____________________________________________________________________________Index S V Serial sensor interfaces CL31, CT25K and CT25KAM 114 GARMIN GPS35 121 introduction 112 PTB220 116 PWD11, PWD21 and FD12P 112 WS425 118 WT500 series 119 Serial to Analog Converter See Nokeval Setup memory size 21 Shaft encoder See Absolute Shaft Encoder Vaisala HUMICAP Humidity and Temperature Transmitters introduction 72 Vaisala Weather Transmitter configuring 56 introduction 55 using sensor variables in reports 61 Va