Manual

Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 39
Output Reading Once
Syntax in STOP-mode: SEND<cr>
Syntax in POLL-mode: SEND aa<cr>
Where
aa
=
Address of the transmitter when more than one transmitter is
connected to a serial bus (0 ... 99).
Example:
>send
aw= 0.2 aw T= -47.37 'C …
If the value is too long to fit to the allocated space or if there is an error
in outputting, the value is displayed with stars ‘*’.
For example,
aw=*.* aw T= 31.0 'C
The output mode can be changed with the FORM command.
Set Serial Interface Mode
Syntax: SMODE x<cr>
Where
x
=
STOP/RUN/POLL
STOP
mode: Transmitter in standby for serial
commands
RUN
mode: Transmitter outputs data continuously
POLL
mode: Transmitter only responds to
addressed command
s. Use the OPEN <transmitter
address>
command to temporarily enable serial
commands
(as in STOP mode) when in POLL
mode.