Example:
>smode run
Output mode : RUN
>r
02:08:01.03 aw= 0.2 aw T= -47.13 'C …
02:08:04.21 aw= 0.2 aw T= -47.16 'C …
<ESC>
>smode stop
Output mode : STOP
Serial Line Settings
Syntax: SERI b p d s<cr>
Where
b
=
bauds (300, 600, 1200, 2400, 4800, 9600,19200)
p
=
parity (n = none, e = even, o = odd)
d
=
data bits (7 or 8)
s
=
stop bits (1 or 2)
The settings can be changed one parameter at a time or all parameters at
once.
Example:
>seri o changing parity only
4800 o 7 1
>seri 600 n 8 1
changing all parameters
600 n 8 1
Set Transmitter Address for Use in POLL
Mode
Syntax: ADDR aa<cr>
Where
aa
=
Address (0 ... 99)
Example:
>addr
Address : 0
>addr 1
Address : 1