Revert Factory Calibration
Syntax: LI<cr>
1. Remove the transmitter unit from the mounting (see page 18) and
press the adjustment button once (see Figure 15 on page 56).
2. Issue the LI command and set the offset value to 0 and gain value
to 1.
>li
RH offset : -0.6000000 ? 0
RH gain : 1.00000000 ? 1
T offset : 0.00000000 ? 0
T gain : 0.40000000 ? 1
>
3. Reset the transmitter by giving the RESET command.
The transmitter returns to normal mode.
View Current Offset and Gain Adjustment
View the currently applied standard user adjustment offset and gain with
the L command.
Syntax: L<cr>
Example:
>l
RH offset : 0.00000000
RH gain : 1.00000000
T offset : 0.00000000
T gain : 1.00000000
>